AXIOMET AX-C708 [3/13] Ranges
![AXIOMET AX-C708 [3/13] Ranges](/views2/1664989/page3/bg3.png)
5. Ranges
A measurement range determines the highest value and resolution at which the meter can measure. Most
meter measurement functions have more than one range (see “Specifications”).
Make sure the correct range is selected:
- If the range is too low, the display shows OL (overload).
- If the range is too high, the meter will not be displaying its most precise measurement.
The meter normally selects the lowest range that will measure the applied input signal (<AUTO> showing
on the display). Press <RANGE> to lock the range. Each time <RANGE> is pressed, the meter selects the
next higher range for every press .At the highest range, it returns to the lowest range.
6. Testing Diodes
To test a single diode:
1. Insert the red test lead into the Hi (MEASURE) jack and black test lead into the COM jack.
2. Set the rotary function switch to Ω .
3. Press the blue button so that the symbol is on the display.
4. Touch the red probe to the anode and the black probe to the cathode (side with band or
bands).The meter should indicate the appropriate diode voltage drop.
5. Reverse the probes. The meter displays OL, indicating a high impedance.
6. The diode is good if it passes the tests in steps 4 and 5.
7. Using Display Hold
Warning
To avoid possible electric shock, do not use Hold to determine if dangerous voltage is present.
Activate Hold to freeze the meter’s display on each new stable reading. Press <HOLD> to activate Hold.
8. Using Relative Measuring Function
Press <REL> , the meter would show the present value in the sub-display district as a reference and the
relative value (minus between present value and referenced value) in the main-display district. For next
press <REL>, the meter would stop displaying the relative value.
Warning
In REL mode, be careful for possible dangerous voltage.
Содержание
- Ax c708 1
- Introduction 1
- Safety information 1
- Input impedance 2
- Measuring electrical parameters 2
- Ranges 3
- Testing diodes 3
- Using display hold 3
- Using relative measuring function 3
- Dc voltage output 4
- Frequency output 4
- Measuring frequency duty cycle ratio 4
- Output function 4
- Dc current output 5
- Simulating transmitter output absorption current 5
- Power up options 6
- Replacing the batteries 6
- Replacing a fuse 7
- Technical specifications 7
- Dc mv measurement 8
- Dc voltage measurement 8
- Ac voltage measurement 9
- Dc current measurement 9
- Ac current measurement 10
- Resistance measurement 10
- Basic technical specification of output 11
- Diode test and continuity test 11
- Frequency count accuracy 11
- General specifications 12
- Appendix 13
Похожие устройства
- PEAKTECH PKT-P4390 Инструкция по эксплуатации
- AXIOMET AX-C850 Расширенная инструкция
- AXIOMET AX-C850 Инструкция
- AXIOMET AX-C830 Расширенная инструкция
- AXIOMET AX-C830 Инструкция
- AXIOMET AX-C800 Расширенная инструкция
- AXIOMET AX-C800 Инструкция
- AXIOMET AX-C705 Расширенная инструкция
- AXIOMET AX-C705 Инструкция
- AXIOMET AX-C605 Расширенная инструкция
- AXIOMET AX-C605 Инструкция
- AXIOMET AX-6003P Инструкция
- AXIOMET AX-6003P Расширенная инструкция
- AXIOMET AX-3005PQ Инструкция
- AXIOMET AX-3005PQ Расширенная инструкция
- AXIOMET AX-3003P Инструкция
- AXIOMET AX-3003P Расширенная инструкция
- AXIOMET AX-3020L Инструкция
- AXIOMET AX-3020L Расширенная инструкция
- AXIOMET AX-3010H Инструкция