AXIOMET AX-C708 [4/13] Output function

AXIOMET AX-C708 [4/13] Output function
9. Measuring Frequency, Duty Cycle Ratio
Press <Hz> to show Frequency, Duty Cycle Ratio and Voltage (Current) in sequence when measuring
voltage (DC mV voltage excluded) and current.
10. Output Function
The output terminal of the meter produces DC voltage, frequency, current and simulating transmitter set
by the user.
Caution!
Using: Do not append voltage to the output jack; otherwise the internal circuit would be damaged for
impropriety.
11. DC Voltage Output
Insert one end of the test lead into the Hi (OUTPUT) jack and COM jack of the meter and connect the
other end to the input of the user’s instrument.
Set the rotary function switch to TCmV (or V ), and press blue button to select mV (or V) function
and displays unit “mV” (or “V”).
Press <LEFT> or <RIGHT> to select the set digits for output.
Press <UP> or <DOWN> to change the numerical value of the set digits. The value can do carry or
number decrement automatically. Hold the pressed key in one second and the value will keep varying.
Press <HOLD> to turn on/off the output followed by displaying the symbol “ON” or “OFF”.
12. Frequency Output
Insert one end of the test lead into the Hi (OUTPUT) jack and COM jack of the meter and connect the
other end to the input of the user’s instrument.
Set the rotary function switch to V , and press the blue button to select function and displays unit
“Hz” indicating frequency set.
Press <LEFT> or <RIGHT> and or <UP> to <DOWN> select output frequency within 1-100Hz.
Press <HOLD> and “ON” is on display, and begin output frequency signal.
Press <RANGE> and displays unit “kHz”, and repeat the 3-4 step to output 0.1-1.1kHz (1.0 - 11.0kHz)
frequency signal.