Fluke FLK-I410 [7/9] Specifications
![Fluke FLK-I1010-KIT [7/9] Specifications](/views2/1667679/page7/bg7.png)
I410/i1010
English
Panel 7
Specifications
Operating Temperature -10 °C to +50 °C (14 °F to 122 °F)
Storage Temperature -20 °C to +60 °C (-4 °F to +140 °F)
Relative Humidity 0 % to 95 % (0 °C to 30 °C)
0 % to 75 % (30 °C to 40 °C)
0 % to 45 % (40 °C to 50 °C)
Altitude
Operating
Non-operating
0 m – 2000 m
0 m – 12000 m
Battery Type 9 V (alkaline) NEDA 1604 IEC 6F22
Battery Life 60 hrs typical (continuous with alkaline)
Safety
s Double Insulation, 600 V rms, CAT. III
Current x Hz Product ≤ 240,000
Output Signal 1 mV per amp dc or ac
Working Voltage 600 V rms, CAT. III maximum at input
Maximum Conductor Size 1 ea. 30 mm (1.18 in.) diameter
2 ea. 25 mm (0.98 in.) diameter
Load Impedance ≥1 MΩ, ≤100 pF
Temperature Coefficient +/-(0.05 x accuracy per °C (0 °C – 18 °C,
28 °C – 50 °C)
Size 209 mm x 78 mm x 48 mm
(3.09 in. x 8.21 in. x 1.87 in.)
Weight 0.5 kg (17.6 oz.)
i410 i1010
Specified Current
Range:
1 A – 400 A ac rms *
1 A – 400 A dc
1 A – 600 A ac rms *
1 A – 1000 A dc
Usable Current
Range:
0.5 A – 400 A 0.5 A – 1000 A
DC Accuracy (zero
adjusted, conductor
centered)
3.5 % + 0.5 A (0 A –
400 A)
2.0 % + 0.5 A (0 A –
1000 A)
AC Accuracy
3.5 % + 0.5 A, 45 Hz –
400 Hz, Crest Factor
≤ 3.
(0 A – 400 A)
2.0 % + 0.5 A, 45 Hz
– 400 Hz, Crest
Factor ≤ 3.
3.0 % + 0.5 A, 400 Hz
- 2 kHz sine wave.
(0 A – 600 A)
Bandwidth 3 kHz 10 kHz
* With a true-rms voltmeter, the minimum ac current is limited to the low
end of the specified mV ac range.
Содержание
- Ac dc current clamp 1
- I410 i1010 1
- Instruction sheet 1
- Safety information 1
- Features and connections 2
- I410 i1010 english panel 2 2
- I410 i1010 english panel 3 3
- Measuring dc current 3
- A connect to voltmeter 4
- B select mv ac or v ac but resolution may be limited to 1 a 4
- C set on 4
- D clamp and center around conductor 4
- E read voltmeter 1 mv 1 a 4
- I410 i1010 english panel 4 4
- Maximum 400 a ac rms i410 or 600 a ac rms i1010 4
- Measuring ac current 4
- I410 i1010 english panel 5 5
- Replacing the battery 5
- Testing the battery 5
- Cleaning 6
- Days remove and store the battery separately 6
- I410 i1010 english panel 6 6
- If the current clamp does not work 6
- Periodically wipe the case with a damp cloth and mild detergent do not use abrasives or solvents 6
- Service and parts 6
- Storage 6
- Symbols 6
- The current clamp should be serviced only by a qualified service technician for service information contact your nearest fluke dealer or service center 6
- I410 i1010 english panel 7 7
- Specifications 7
- I410 i1010 english panel 8 8
Похожие устройства
- Fluke FLK-I410 Инструкция язык en
- Fluke FLK-I410-KIT Инструкция по эксплуатации
- Fluke FLK-I410-KIT Расширенная инструкция
- Fluke FLK-I410-KIT Инструкция язык en
- Fluke FLK-I30 Инструкция язык en
- Fluke FLK-I30 Расширенная инструкция
- Fluke FLK-I30 Инструкция по эксплуатации
- Fluke FLK-I30S Расширенная инструкция
- Fluke FLK-I30S Инструкция язык en
- Fluke FLK-I30S Инструкция по эксплуатации
- Fluke FLK-I30S Data Sheet
- Fluke FLK-I310S Инструкция язык en
- Fluke FLK-I310S Инструкция по эксплуатации
- Fluke FLK-I310S Расширенная инструкция
- Gigabyte X470 AORUS GAMING 7 WIFI (rev. 1.0) Инструкция
- Gigabyte X470 AORUS GAMING 5 WIFI (rev. 1.0) Инструкция
- Gigabyte X570 I AORUS PRO WIFI Инструкция
- Gigabyte X570 AORUS XTREME (rev. 1.0) Инструкция
- Gigabyte X570 AORUS ULTRA (rev. 1.1/1.2) Инструкция
- Gigabyte X570 AORUS ULTRA (rev. 1.0) Инструкция