FAULTS
FAULTS REASON
SOLUTION
The device doesnt
work
Turn off very quickly
The device turn off
during the working
mode
Incorrect polarity of power
supply units installation
Electrodes installed uncorrectly
The massage finished because
timers ends after 15 min
Install the power supplies while
maintaining polarity
Put electrode's stickers correctly
Install new battaries type AA, or
conect to socket the power cord
Power cord/adapter is broken
Weak or discharged power
supplies
The adhesive
surfaces of the
plates are too
adhesive
Plate applied to skin immediately
after washing
plate adhesion surface is
damaged
Transparent plastic film not
removed
Electrode stickers
too much sticky
Electrodes washed too long
or often
Electrodes was storaged on high
temperatures or directly under
sunchine
Electrodes used not on dry body
Feeling of
stumulation
effect is poor
Electrode overlaps one another
Bad Contact , in the place where
power cord connected
Electrodes installed badly
Low installed intensity
Redness or
irritation of the skin
when stimulated
Damage of electrode's surface
Dry or dirty surface of
electrodes
Try to make massage again or
turn off the device
Change the power cord or batteries AA
Remove the film from the adhesive
surface of the electronic plate
Dry the electrodes
Change the electrodes
Put electrodes in the frige
for a night
Put electrodes densely
Apply electrodes to the body
without overlapping
Check again good connection
with powercord
Increase the intensity of
stimulation, using channel 1 or 2
Wash the adhesive surfaces of the
plates for 3 seconds with the fingertips
under a weak stream of water
Change electrode - stickers
41
www.planta.ru.com