Graco 255277 LDP5 [12/34] Calibration

Graco 24F886 LDP5 [12/34] Calibration
LDM5 Meter Setup and Operation Instructions
12 312668Z
Calibration (FIG. 13)
Recalibrates the meter for dispensing different fluids.
1. If the Calibration Menu is not already displayed,
complete steps 1 - 4 of the Setup Menus section,
page 10.
2. CAL blinks indicating you are on the Calibration
Menu screen. Do ONE of the following.
Use the current calibration.
Press and hold Reset but-
ton to lock in amount. Dis-
play returns to the
Standard Mode.
OR
Recalibrate meter as follows:
a. If the unit of measurement is gallons, pints,
or quarts, dispense exactly one quart of fluid
into a calibrated 1–quart container.
If the unit of measurement is liters, dispense
exactly one liter of fluid into a calibrated
1–liter container.
For proper calibration, you must dis-
pense the exact amount.
b. Press and hold Total
button until CAL stops
blinking and the dis-
play shows 1.00.
c. When CAL starts to
blink again, the display
should show 1.00. The
new calibration is complete.
OR
To return to the default calibration:
Press and hold the Total button until CAL
stops blinking and the display shows 0.98 (if
the unit of measurement is gallons, pints or
quarts) or 0.99 (if the unit of measurement is
liters).
3. Press and hold the Reset but-
ton. The unit returns to Stan-
dard Dispense Mode.
F
IG. 13
GAL
QTSPTS
LITERS
R-TOTAL
CAL
Min. Dispense Volume 0.5 liter (0.13 gal)
Minimum Flow Rate 1.0 l/min. (0.26 gpm)
Maximum Flow Rate 19.0 l/min. (5.00 gpm)
Maximum Working Pressure: 70 bar (1000 PSI)
ti12046a
If an error was made during meter recalibration,
repeat Steps a - c of the recalibration process to
recalibrate the meter again.