6. The Laser indicates the approximate area of detection.
7. LCD displays the Dew Point value after 2 seconds even though other
values are still drifting under stabilization mode.
8. Release the Activation Trigger. LCD holds all temperature and humidity
display readings including dew bar.
9. Timer count down 2 minutes if no key is pressed. Auto power off.
STABILIZATION MODE
1. While taking measurement, if the ambient temperature and relative
humidity readings are NOT stable or more than +/- 1 reading units the
stabilization icon displays with bar appearing sequentially. Once its thermal
and hydro equilibrium reached within +/-1 units (e.g. +/-1 deg.C or +/- 1% RH
stable) the stabilization icon disappears.
MAX / MIN / AVG
1. Users can read the Maximum, Minimum, and Average of the last
temperature and humidity reading value by pressing the Max/Min/Avg
button in cycle.
2. Users can delete (clear) the stored value of Maximum, Minimum and
Average surface temperature by pressing and holding the Max/Min/Avg
button for 2 seconds.
EMISSIVITY MODE
1. Press “Max/Min/Avgkey repeatedly until it reach the Emissivity setting
mode.
2. The default value of emissivity value is 0.95
3. Press Activation Trigger to adjust the emissivity value.
4. In the Emissivity mode, the IR temperature is shown for reference.
IR Temp
Flashing
Emissivity