6 Using the software
20
6.3.2. Parameter
The settings from the recording of the thermal image can be
corrected.
Changes to the settings influence the measurement
result. The settings should therefore only be changed
with the utmost caution.
Function Procedure
Set Emissivity,
Reflected temperature,
Intensity,
Humidity
1
,
Ambient temperature
1
> Click on the number and
change it by means of input via
the keyboard or via .
You can find more information on
the individual properties in the
instruction manual of the thermal
imager.
Dewpoint Dewpoint temperature: The value
is calculated from the humidity and
the ambient temperature and
cannot be changed manually.
Discard changes > If you wish to reset the
changes back to the most
recently saved parameters:
Click on Discard changes .
6.3.3. Playing/saving audio comments
The function is only available if an audio comment is attached to
the thermal image currently selected (see status bar). It is not
possible to record, change or delete audio comments.
> Click on the control key symbols to control the audio functions:
(Rewind), (Play), (Pause), (Stop), (Fast-
forward).
> Click on to save the audio file in the file format MP3 or
Wave.
1
The humidity and ambient temperature are transmitted by the thermal imager
and can be changed in IRSoft.
When shown as a humidity image: correct values must be entered to calculate
the relative surface moisture. When shown as a temperature image: the values
are for information purposes only.