Fluke 773 [9/26] Introduction

1
Introduction
XW Warning
To prevent electrical shock or personal injury, do not do the
calibration verification tests or calibration procedures in this
manual unless you are qualified.
The data in this manual is for qualified personnel only.
This manual tells you about verification and adjustment procedures for the 772/773
Milliamp Process Clamp Meter (referred to in this manual as the Meter). The Meter
features closed-case calibration to use with reference sources. It measures the reference
signals, calculates the correction factors, and keeps them in memory. Calibration
adjustment is required after a repair, or if the Meter fails a performance test.
This manual explains:
Precautions and Safety Information
Specifications
Basic Maintenance
Calibration/Verification Procedure
Replaceable Parts and Accessories
For complete use instructions, refer to the 772/773 Instruction Sheet.