Fluke 754: The HART calibrator that is easy to use.
HART applications
Versatile HART protocol support
The 754 supports the commands contained in HART protocol version 5.7.
The 754 supports a substantial set of HART instructions:
Universal commands—provide functions that are implemented in all
field devices, for example, read manufacturer and device type, read
primary variable (PV), or read current output and percent of span
Common practice commands—provide functions that are common to
many but not all field devices, for example read multiple variables, set
damping time, or perform loop test
Device-specific commands—provide functions that are unique to a
particular field device, for example sensor trim. The 754 supports these
devices:
Todays 754 supports device-specific instructions for a variety of popular
instruments. Additional support may be added periodically with a simple
software update available on diskette or via download for a modest
upgrade fee.
Manufacturer Pressure
instruments
Temperature
instruments
Coriolis
instruments
ABB/Kent-Taylor 600T 658T
1
ABB/Hartmann &
Braun
Contrans P,
1
AS 800 Series
Endress & Hauser CERABAR S, CERABAR M,
DELTABAR S
TMT 122
1
, TMT 182
1
,
TMT 162
1
Foxboro Eckardt TI/RTT20
1
Foxboro/Invensys I/A Pressure
Fuji FCX, FCXAZ FRC
Honeywell ST3000 STT25T
1
, STT25H
Micro Motion 2000, 2000 IS,
9701, 9712, 9739
Moore Products 344
1
Rosemount 1151, 2088, 3001C,
3051, 3051S
3044C, 644, 3144,
3244, 3144P
Siemens SITRANS P DS, SITRANS
P ES
SMAR LD301 TT301
1
Viatran I/A Pressure
Wika UNITRANS T32H
1
Yokogawa EJA YTA 110, 310, 320
www.uke.com/processtools
1
Sensor Trim not supported
5 Fluke Corporation 750 Series Brochure
HART operating
modes supported
For Point to Point operation, the
most commonly used mode, con-
nects the 754 to a single HART
device in a 4-20 mA loop.
In Multi-Drop mode, several
HART instruments can be bussed
together. The 754 searches for
each, identifies addresses in
use, and allows you to select the
instrument for calibration and
related operations.
In Burst Mode, the HART instru-
ment transmits bursts of data
without waiting to be interro-
gated by a master unit. The 754
can take transmitters out of burst
mode during test or calibration,
then later restore them to burst
mode.