Error codes
IKA
®
Rocker 3D basic
In case of a fault, a beep is heard and the “Power” LED flashes.
Proceed as follows in such cases:
Switch the device off.
Perform corrective measures as described below.
Perform a restart.
Error Cause Effect Solution
Motor blockage or overload Motor off - Switch the device off.
- Reduce the load on the motor and check all moving parts for
freedom of movement.
- Then perform a restart.
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the IKA
®
service department,
- Send the device for repair, including a short description of the fault.
IKA
®
Rocker 3D digital
The fault is shown by an error code in the display as following if the error occurs, e.g. Er 4.
Proceed as follows in such cases:
Switch the device off.
Perform corrective measures as described below.
Perform a restart.
Error Cause Effect Solution
Er 3 Temperature inside device is too
high
Motor off - Switch the device off.
- Allow the device to cool down.
Er 4 Motor blockage or overload Motor off - Switch the device off.
- Reduce the load on the motor and check all moving parts for
freedom of movement.
- Then perform a restart.
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the IKA
®
service department,
- Send the device for repair, including a short description of the fault.