712 English Instruction Sheet
Page
5
Simulating an RTD
To simulate an RTD, proceed as follows:
1. Press O to turn on the power.
2. If the calibrator is in input mode (INPUT on the display),
press INPUT OUTPUT once. Make sure the display shows
OUTPUT.
3. Press
R to select the desired RTD type.
4. Connect test leads to the terminals of the RTD measuring
device as shown. Use only the two center outputs (labeled
2W 3W 4W OUTPUT).
NC NC
4W 3W
2W 3W 4W
OUTPUT
INPUT
30V
MAX
30V
MAX
30V
MAX
Sense
connection
(if needed)
Sense
connection
(if needed)
RTD
measurement
device input
kg01f.eps