General directions
6,000- and 60,000-count digital multimeters 7
General directions (continued)
Maintenance
Disconnect everything connected to the instrument and press the ON/OFF
key ( ).
Use a soft cloth, moistened with soapy water.
Rinse with a damp cloth and dry rapidly with a dry cloth or forced air.
Make sure that no foreign objects interfere with the operation of the device
by which the leads are snapped into place.
Replacing
the fuse
Before replacing the fuse (reached by
opening the bottom compartment), disconnect
the instrument from any source of current.
During the replacement, make sure that only a
fuse of the appropriate rating and specified
type is used. Using another type of fuse and
shorting the fuse holder are strictly forbidden.
Checking the current fuse:
Fuse: SIBA/5019906
MTX 3291: 11A: 10x38 1,000V - F
breaking capacity: >20kA
MTX 3290: 10A: 6x32 - 600V - F
breaking capacity: >50kA
Rechargeable
and primary batteries
The multimeter is powered by primary or rechargeable batteries (see above).
To charge the rechargeable
batteries (set of 4 NI-MH LSD
batteries), use an external rapid
charger, available as an accessory.
After replacing the batteries,
wait 10s before switching the instrument back on.
Active
communication
interface
(MTX 3291 only)
The multimeter can communicate with a PC via the USB link.
The basic version includes a USB link using an isolated optical USB cord (type
HX0056Z) and SX-DMM software, plus Labview and Labwindows drivers to
program the devices.
MTX 3291: They can also be programmed via the SCPI protocol:
- to program via Labview/LW
- to recover data or program the instrument using the software
- to calibrate the MTX 3291