Do not use the instrument in an atmosphere which
contains sulfuric acid mist or other substances which
cause corrosion to metal.
Do not use the instrument in a dusty place or a highly
humid place as such will cause instrument reliability
degradation and instrument failures.
Install the instrument in a place where is free from
vibration.
Install the instrument in a place where the ambient
temperature is in range of 10~70°C. Note that the
instrument operation may become unstable if it is
operated in an ambient temperature exceeding the range
of 0~40°C
2.3. Power Supply
AC Input voltage: 110V~120V/220~240V ±10%, 50/60Hz.
Connect the protective grounding conductor of the AC power
cord to an earth ground to avoid electrical shock.
2.4. Fuse
Make sure the correct type of fuse is installed before
power up.
Replace the AC fuse with the same type and rating as
the original fuse.
Disconnect the power cord before fuse replacement.
Make sure the cause of fuse blowout is fixed before
fuse replacement.
2.5. Cleaning
Before cleaning, disconnect the AC mains.
To clean the power supply, use a soft cloth dampened in
a solution of mild detergent and water. Do not spray
cleaner directly onto the instrument, since it may leak
into the cabinet and cause damage.
Do not use chemicals containing benzene, benzene,
toluene, xylene, acetone, or similar solvents.
Do not use abrasive cleaners on any portion of the
instrument.
2.6. Operation notes
EN 61010-1:2001 specifies the pollution degrees and their
requirements as follows. The instrument falls under degree
2.
3. Quick Starting
3.1. Front Panel and Rear Panel
Front Panel