Beurer FT70 [29/76] Technical data 11

Beurer FT70 [29/76] Technical data 11
29
Technical data 11
Tip: If the device is used other than in accordance with the specifications, perfect functioning cannot be
guaranteed. We reserve the right to make technical changes to improve and develop the product.
This device complies with European Standard EN60601-1-2 and is subject to particular precautions with re-
gard to electromagnetic compatibility. Please note that portable and mobile HF communication systems may
interfere with this device. For more details, please contact Customer Services at the address indicated.
Name and model
SFT65
Measurement range Ear/Forehead thermometer mode: 93.2°F – 109.4°F (34°C - 43°C)
Object temperature mode: 32°F – 212°F (0°C - 100°C)
Laboratory
measurement
precision
Ear thermometer mode: ±0.4°F (±0.2°C) from 95.9°F - 107.6°F (35.5°C - 42°C),
outside of this measurement range ±0.5°F (±0.3°C)
Forehead thermometer mode: ±0.5°F (±0.3°C) from 71.6°F - 104°F (22°C - 40°C)
Object temperature mode: ±2.7°F (±1.5°C) at < 86°F (<30°C);
±5% at ≥ 86°F (≥ 30°C)
Interval between
2 measurements
At least 5 seconds
Clinical repeat
precision
Ear: Children, 1 – 5 years: ±0.14°F (±0.08°C)
Adults: ±0.13°F (±0.07°C)
Forehead: Children, 1 – 5 years: ±0.13 °F (±0.07°C)
Adults: ±0.14°F (±0.08°C)
Measurement units °Celsius (°C) or °Fahrenheit (°F)
Ambient operating
conditions
60.8°F - 95°F (16°C - 35°C) with a relative humidity of up to 85% (non-condensing)
Storage conditions -13°F to 131°F (-25°C - 55°C) with a relative humidity of up to 85%
(non-condensing)

Содержание