Beurer FT70 [30/76] Trouble shooting 12

Beurer FT70 [30/76] Trouble shooting 12
30
Dimensions 38.2 x 138 x 46.5 mm
Weight 90 g including battery
Battery 2x AAA, LR 03
Memory For 10 measurements
Symbol key
Device classification type BF
Please read the instructions for use.
Trouble-shooting12
Display Cause Solution
The temperature measured is higher than
1) Ear/Forehead thermometer mode: 109.4°F (43°C)
2) Object temperature mode: 212°F (100°C)
Operate the thermometer only between the
specified temperature ranges. If necessary,
clean the sensor tip. In the event of a
repeated error message, contact your
retailer or Customer Services.
The temperature measured is lower than
1) Ear/Forehead thermometer mode: 93.2°F (34°C)
2) Object temperature mode: 32°F (0°C)
The operating temperature is not in the range from
60.8°F - 95°F (16°C - 35°C)
Operate the thermometer only between the
specified temperature ranges.