B.Well WA-77 [16/64] Instruction manual wa 77

B.Well WA-77 [16/64] Instruction manual wa 77
Instruction manual
WA-77
16
Eng
Error No. Possible cause(s)
Err 1 Pulse signals on cu are too weak, no pulse has been detected.
Err 2
Unnatural pressure impulses in uence the measurement result. Reason: The arm was moved during the Measure-
ment (Artefact).
Err 3 In ation of the cu takes too long. The cu is not correctly seated.
ERR 5
The measured readings indicated an unacceptable di erence between systolic and diastolic pressures. Take another
measurement following directions carefully. Contact you doctor if you continue to get unusual readings.
«LO» The pulse is too low (less than 40 beats per minute). Repeat the measurement.
«HI»
The pressure in the cu is too high (over 300 mmHg) OR the pulse is too high (over 200 beats per minute). Relax for
5 minutes and repeat the measurement.