709/709H/710
Calibration Manual
8
3. Make sure that the Product shows SOURCE in the upper-left corner.
4. Set the DMM to measure dc current.
5. See Table 2 to verify the readings on the DMM for the subsequent tests. (No adjustment
is necessary for Test 1.)
6. Test 2: Push twice to select 12.000 mA. Verify the readings.
7. Test 3: Push twice to select 20.000 mA. Push and rotate the knob until 24.000 mA is
shown. Verify the readings.
Table 2. DC Current Source, Current Measure, Voltage Measure Mode Tests
Test
Number
Product Output Mode
8508A DMM
Minimum
8508A DMM
Maximum
1
2
3
4.000 mA
12.000 mA
24.000 mA
mA Source
mA Source
mA Source
3.9976 mA
11.9968 mA
23.9956 mA
4.0024 mA
12.0032 mA
24.0044 mA
Test
Number
5522A Calibrator
Output
Mode
Product
Minimum
Product
Maximum
4
5
6
4.000 mA
12.000 mA
24.000 mA
mA Measure
mA Measure
mA Measure
3.998 mA
11.997 mA
23.996 mA
4.002 mA
12.003 mA
24.004 mA
Test
Number
5522A Calibrator
Output
Mode
Product
Minimum
Product
Maximum
7
8
9
0.000 V
14.000 V
28.000 V
Volt Measure
Volt Measure
Volt Measure
-0.002 V
13.997 V
27.995 V
0.002 V
14.003 V
28.005 V
DC Current Measurement Mode
To verify the performance of the dc current measurement function:
1. Push .
2. Turn the rotary knob to select mA Measure.
3. Push the rotary knob.
The Product shows MEASURE.
4. Disconnect all connections to the Product.
5. Connect the product:
a. Connect a test lead from the red AUX terminal of the DC Calibrator (Calibrator) to the
DMM I input.
b. Connect a test lead from the black AUX terminal of the Calibrator to the COM
terminal on the Product.
c. Connect a test lead from the DMM LO input to the Product + terminal. See Figure 3.
Содержание
- 709h 710 1
- 709h 710 calibration manual 1
- Warranty 2
- Table of contents 3
- Contact fluke 5
- Introduction 5
- To prevent possible electrical shock fire or personal injury read all safety information before you use the product 5
- Xw warning 5
- Safety information 6
- Specifications 6
- Battery replacement 8
- Clean the product 8
- Maintenance 8
- Xw warning 8
- Precision loop calibrator 9
- 709h 710 10
- Performance verification tests 10
- Required test equipment 10
- Table 1 is a list of the required equipment for the performance verification tests and calibration adjustment 10
- The performance tests verify the complete operation of the product and measure the accuracy of each function against product specifications if the product fails a part of the test calibration adjustment or repair is necessary see calibration adjustment procedure 10
- The performance verification tests check the accuracy of each product function against its specifications if the product fails any of these tests calibration adjustment or repair is necessary 10
- To prevent possible electrical shock fire or personal injury do not perform the performance test or verification test procedures unless the product is fully assembled 10
- Xw warning 10
- Dc current source mode 11
- 709h 710 12
- A connect a test lead from the red aux terminal of the dc calibrator calibrator to the dmm i input 12
- B connect a test lead from the black aux terminal of the calibrator to the com terminal on the product 12
- C connect a test lead from the dmm lo input to the product terminal see figure 3 12
- Connect the product 12
- Dc current measurement mode 12
- Disconnect all connections to the product 12
- Make sure that the product shows source in the upper left corner 12
- Push the rotary knob 12
- See table 2 to verify the readings on the dmm for the subsequent tests no adjustment is necessary for test 1 12
- Set the dmm to measure dc current 12
- Test 2 push twice to select 12 00 ma verify the readings 12
- Test 3 push twice to select 20 00 ma push and rotate the knob until 24 00 ma is shown verify the readings 12
- The product shows measure 12
- To verify the performance of the dc current measurement function 12
- Turn the rotary knob to select ma measure 12
- Dc voltage measurement mode 14
- Hart transmitter 709h 710 only 15
- Calibration adjustment procedure 16
- Ma measure mode 16
- Ma source 17
- Voltage measure 18
- Precision loop calibrator 19
- User replaceable parts 19
- User replaceable parts are listed in table 3 and shown in figure 6 for more information about these items and their prices contact a fluke representative see the contact fluke section 19
- 709h 710 20
- Calibration manual 20
- Figure 6 user replaceable parts 20
Похожие устройства
- Fluke 773 Приложение к Инструкции
- Fluke 773 Положение об энергозависимости памяти
- Fluke 773 Clamp Meter data sheet
- Fluke 773 Руководство по калибровке
- Fluke 772 Положение об энергозависимости памяти
- Fluke 772 Приложение к Инструкции
- Fluke 772 Clamp Meter data sheet
- Fluke 772 Руководство по калибровке
- Fluke 771 Приложение к Инструкции
- Fluke 771 Приложение к руководству по калибровке
- Fluke 771 Руководство по калибровке
- Fluke 715 Положение об энергозависимости памяти
- Fluke 715 Loop Calibrators Data Sheet
- Fluke 715 Приложение к Инструкции
- Fluke 718 Приложение к Руководству по эксплуатации
- Fluke 712 Приложение к Инструкции
- Fluke 714 Приложение к Инструкции
- Fluke 717 Приложение к руководству по калибровке
- Fluke 717 Приложение к Инструкции
- Fluke 717 Руководство по калибровке
Скачать
Случайные обсуждения