Fluke 709 [14/20] Dc voltage measurement mode

Fluke 709 [14/20] Dc voltage measurement mode
709/709H/710
Calibration Manual
10
DC Voltage Measurement Mode
To verify the performance of the dc voltage measurement function:
1. Push .
2. Turn the rotary knob to select Volts Measure.
The Product display shows Measure with V next to the measurement.
3. Connect the test leads from the output NORMAL terminals of the Calibrator to the input
terminals on the Product (black to COM and red to [+]). See Figure 4.
0
/
+
5522A
hid012.eps
Figure 4. DC Voltage Measurement Verification Connections
4. Set the Calibrator to the voltage in test 7 in Table 2 and verify the display reading on the
Product.
5. Repeat for tests 8 and 9. The reading on the display should be within the minimum and
maximum values shown in Table 2.
6. Set the Calibrator output to 0 V and set the output to STANDBY.
7. Turn the Product off and disconnect it from the Calibrator.
If the Product failed any of these tests, calibration adjustment or repair is necessary.
The performance verification tests for the 709 are now complete. For the 709H see the
next section.

Содержание

Скачать