712B/714B
Calibration Manual
30
Adjust 0.1 mA External Trigger
1. Set the 5522A to output 0.1 mA.
2. Push on the 712B to enter 0.1 mA external trigger calibration mode.
3. Wait approximately 10 seconds for a stable output of the 5522A.
4. Confirm the reading on the 712B.
hwp16.jpg
Adjust 1 mA External Trigger
1. Set the 5522A to output 1 mA.
2. Push on the 712B to enter 1 mA external trigger calibration mode.
3. Wait approximately 10 seconds for a stable output of the 5522A.
4. Confirm the reading on the 712B.
hwp17.jpg