Fluke 714B [7/50] List of figures
Превью страниц
Страница 7 /
50
v
List of Figures
Figure Title Page
1. Replace the Batteries ................................................................................
9
2.
Verify 712B Ohms Source ......................................................................... 11
3.
Verify 712B 4-Wire Ohms Measure ........................................................... 12
4.
Verify 712B 3-Wire Ohms Measure ........................................................... 13
5.
Verify 712B mA Measure .......................................................................... 15
6.
Verify 714B mV Source ............................................................................. 16
7.
Verify 714B mV Measure .......................................................................... 17
8.
Verify Type K Measure .............................................................................. 18
9.
Verify mA Measure .................................................................................... 20
10.
Adjust 712B Ohms Source ........................................................................ 21
11.
Adjust 712B Trigger Current ...................................................................... 24
12.
Adjust for 712B Ohms Measure ................................................................ 26
13.
Adjust 712B mA Measure .......................................................................... 29
14.
Adjust External Trigger for Ohms Source .................................................. 30
15.
Connect the 714B to the 8508A ................................................................ 33
16.
Adjust 714B mV Measure .......................................................................... 34
17.
Connect the 714B to the Fluke 9101-256 .................................................. 36
18.
Use Lag Bath to Adjust Cold Junction Temperature ................................. 37
19.
714B and 5522A mA Channel Connections .............................................. 38
20.
User-Replaceable Parts and Accessories for 712B .................................. 41
21.
User-Replaceable Parts and Accessories for 714B .................................. 43
Содержание
147- B 714b calibration manual
- B 714b
- Warranty
- Limited warranty and limitation of liability
- Table of contents
- List of tables
- List of figures
- Warning
- Safety information
- Introduction
- Contact fluke
- Symbols
- Ohms measurement
- Dc ma measurement
- B specifications
- Specifications are based on a one year calibration cycle and apply from 18 c to 28 c unless stated otherwise all specifications assume a 5 minute warmup period
- Rtd thermocouple calibrator
- Ohms source
- Rtd input and output
- Calibration manual
- B 714b
- Rtd thermocouple calibrator
- General specifications
- Thermocouple ma input
- Thermocouple input and output
- Specifications are based on a one year calibration cycle and apply from 18 c to 28 c unless stated otherwise all specifications assume a 5 minute warmup period
- Calibration manual
- B specifications
- B 714b
- Accuracy of reading floor
- Accuracy of output floor
- Thermocouple mv output
- Thermocouple mv input
- Rtd thermocouple calibrator
- B specifications
- General specifications
- Calibration manual
- B 714b
- Warning
- Replace the batteries
- Maintenance
- Clean the product
- The equipment listed in table 2 is necessary to do the calibration and test for the 712b the equipment listed in table 3 is necessary to do the calibration and test for the 714b
- Required equipment
- Note fluke 5522a and fluke 8508a need an 8 hour warmup period before the calibration starts the 5522a needs to zero all the functions weekly and zero ohms output function every day the 8508a needs to zero the dc v measurement function the 4 wire ohms measurement function and the dci measurement function
- B 714b
- B performance verification
- Verify ohms source
- Verify 4 wire ohms measure
- Set the 5522a to output the ohms value at the second column in table 5 5 wait approximately 10 seconds for a stable output of the 5522a and verify that the 712b readings are within the limits shown
- Push until 4w shows on the screen
- Push to set the 712b to ohms measure mode
- Make the connections between the 712b and the 5522a as shown in figure 3
- B 714b
- Wait approximately 10 seconds for a stable output of the 5522a and verify that the 712b readings are within the limits shown
- Verify 3 wire ohms measure
- Until 3w shows on the screen
- To set the 712b to ohms measure mode
- Set the 5522a to
- Output the ohms value at the second colum
- Make the connections between the 712b and the 5522a as shown in figure
- In table 6
- Adjust the ma source output of 5522a and make that 8508a shows as the first column in the table 7 and verify that the 712b readings are within the limits shown
- Verify ma measure
- Set the 8508a to dci measurement mode set the range of the dci measurement mode to 20 ma and the resolution of dci measurement mode to 7 1 2 digits
- Make the connections between the 712b 8508a and 5522a as shown in figure 5
- B 714b
- Rtd thermocouple calibrator
- Turn the 714b on and let it warm up for 5 minutes
- Set the 714b to source mv mode
- Set the 714b to output the mv values in table 8
- Make connections between the 714b and the 8508a with copper wire as shown in figure 6
- D push as necessary to enter source mode
- Check that the 8508a readings are within the appropriate limits in table 8
- C push again
- B push until mv is highlighted
- B performance verification
- B 714b
- A push
- Verify that the 714b performance is within the 1 or 2 year specifications as desired
- Verify mv source
- Verify tc measure type k
- Verify mv measure
- Use k type tc sensor to connect the 7526a and the 714b as shown in figure 8
- Set the 7526a to tc out type k
- Set the 714b to type k measure mode
- Rtd thermocouple calibrator
- Push to set the 714b to mv measure mode 3 set the 5522a to the first value in table 9 4 output the mv values in table 9 and verify that the 714b readings are within the limits shown
- Make connections between the 714b and the 5522a with copper wire as shown in figure 7
- Set the 7526a to the values in table 10 and verify that the 714b readings are within the limits shown
- Set the 7526a to tc in type k
- Set the 714b to type k source mode
- Set the 714b to the values in table 11 and verify that the 7526a readings are within the limits shown
- Maintain the connections in figure 8
- B 714b
- Verify tc k source
- Verify ma measure
- Rtd thermocouple calibrator
- Connect the 5522a 8508a and 714b as shown in figure 9 2 set the 8508a to dci measurement mode set the range of the dci measurement mode to 20 ma and the resolution of dci measurement mode to 7 1 2 digits 3 set the 5522a to source and make sure that 8508a shows the first dc current in table 12 4 output the dc current values in table 12 and verify that the 714b readings are within the limits shown
- Enter manual calibration mode
- B calibration adjustment
- Adjust ohms source
- Adjust 50 ω source
- Adjust 350 ω source
- Adjust 500 ω source
- Adjust 3500 ω source
- Adjust trigger current for ohms measure
- Adjust 0 ma trigger current
- Adjust 0 ma trigger current
- Adjust 0 5 ma trigger current
- Adjust 1 ma trigger current
- Adjust ohms measure
- Adjust 50 ω input
- Adjust 350 ω input
- Adjust 500 ω input
- Adjust 3500 ω input
- Adjust ma measure
- Adjust 0 ma input
- Adjust external trigger for ohms source
- Adjust 24 ma input
- Adjust 1 ma external trigger
- Adjust 0 ma external trigger
- Save data and reset the 712b system
- Enter calibration mode
- B calibration adjustment
- Use 8508a to adjust the 5522a output
- Adjust mv source
- Adjust mv measure
- Use the fluke 9101 256
- Adjust internal cold junction temperature
- Use a lag bath
- The 5522a outputs 0 ma
- Connect the 5522a and the 714b via the fluke 75x 8014 cable as shown in figure 19
- Adjust ma measure
- Save data and reset the 714b system
- Warning
- User replaceable parts for the 712b are listed in table 13 and shown in figure 20 user replaceable parts for the 714b are listed in table 14 and shown in figure 21 for more information about these items contact a fluke representative see the contact fluke section of this manual
- User replaceable parts and accessories
- To prevent possible electrical shock fire or personal injury use only specified replacement parts
- Rtd thermocouple calibrator
- Figure 20 user replaceable parts and accessories for 712b
- Calibration manual
- B 714b
- Rtd thermocouple calibrator
- Figure 21 user replaceable parts and accessories for 714b
- Calibration manual
- B 714b
Похожие устройства
-
Fluke 9144Руководство по эксплуатации -
Fluke 9143Руководство по эксплуатации -
Fluke 9142Руководство по эксплуатации -
Fluke 712CКраткий справочник -
Fluke 712CРуководство по эксплуатации -
Fluke 712CПриложение к Руководству по эксплуатации -
Fluke 712CПаспорт безопасности -
Fluke 712BРуководство по калибровке -
Fluke 712BКраткий справочник -
Fluke 712BРуководство по эксплуатации -
Fluke 712BПриложение к паспорту безопасности -
Fluke 712BПаспорт безопасности