Beurer IH 50 [17/76] Do not touch the unit with wet hands while it is

Beurer IH 50 [17/76] Do not touch the unit with wet hands while it is
17
Disassembly
To continue with cleaning
its individual components,
disassemble the atomiser
(2, 3):
Remove the medicine
container (2) and mesh
atomiser (3) by pressing
both brackets inwards
and drawing the unit
out.
Remove the mesh
atomiser (3) from the
medicine container (2)
by turning it
anticlockwise and
drawing it out.
The nebuliser is later
reassembled in the re-
verse order.
Important
Never hold the complete device under running wa-
ter to clean it.
When cleaning, ensure that all residues are removed.
Never use any substances that may be poisonous
when in contact with the skin or mucous membranes,
or when swallowed or inhaled.
If required, you can clean the device casing with a
lightly moistened cloth that has been soaked in mild
soapy water.
If any traces of medical solutions or impurities re-
main on the gold-coloured contacts on the device
and atomiser, use cotton buds dampened with ethyl
alcohol to remove them.
Clean the outside of the mesh and outside of the
medicine container with a cotton bud dampened with
ethyl alcohol.
Important
Before you clean the unit, always switch it off, unplug
it and let it cool off.
Do not use abrasive detergents and never immerse
the unit in water.
Ensure that no water penetrates inside the unit!
Do not put the device or accessories in a dishwasher.
Do not touch the unit with wet hands while it is
plugged in; do not allow any water to be sprayed
onto the unit. The unit must be operated only when
it is completely dry.
Do not spray any liquids into the ventilation slots. Any
liquid that penetrates into the unit can damage the
electrical parts or other components of the nebuliser
and impair the function of the device.
Disinfection
Follow the steps listed below carefully in order to disin-
fect your atomiser and the accessories. It is advisable
to disinfect the individual components after the last
time they are used each day at the latest.
You can disinfect the IH 50 and its accessories in two
different ways, either with ethyl alcohol or in boiling
water.
First clean the atomiser and the accessories as de-
scribed under “Cleaning”. You can then continue with
disinfection.
Important
Ensure that you do not touch the mesh (3), as this
may destroy it.
Disinfection with ethyl alcohol
Pour 8 ml ethyl alcohol into the medicine container
(2). Close the container. Leave the alcohol in the con-
tainer for at least 10 minutes.
For better disinfection, shake the unit gently from
time to time.
Afterwards, pour the ethyl alcohol out of the container.
Repeat this process, using water this time.
Place the medicine container (2) in a position that al-
lows you to apply a few drops of ethyl alcohol to the
mesh (3). Allow the alcohol to work for 10 minutes.
Finally, clean all parts again under running water.
Disinfection with boiling water
The atomiser should be disassembled as described
and, together with the mouthpiece, placed in boiling
hot water for 15 minutes. Avoid the parts coming into
contact with the hot base of the pot.
You can also disinfect the atomiser and the mouth-
piece with a standard vaporiser. Observe the manu-
facturer’s instructions for use for the vaporizer when
doing this.
The atomiser must not be placed in the microwave.
Note
The face masks and air hose must not be placed in
hot water!
We recommend disinfecting the masks using a retail
disinfectant.