6.Handling Errors and Problems
Care and maintenance
6. Handling Errors and Problems
6.1 Error Messages
Error Display Cause Remedy
Movement during
measurement
Repeat measurement. Remain still
and do not talk during
measurement.
Refer to section 3.3.
Air plug disconnected.
Insert the air plug securely.
Refer to section 3.2.
Arm cuff not applied
correctly.
Apply the arm cuff correctly.
Refer to section 3.2.
Clothing is interfering with
the arm cuff.
Remove any clothing interfering
with the arm cuff.
Refer to section 3.2.
Air is leaking from the arm
cuff.
Replace cuff with new one.
Refer to Chapter 8.
The arm cuff was inflated
above 299 mmHg when
inflating the cuff manually.
Remove the arm cuff and take
another measurement.
Refer to section 3.3.
Battery power is low.
Replace all four “AAA” batteries
with new ones.
Refer to section 2.1.
Blinks or appears
continuously
HEM-7301-ITKE_main.book 32 ページ 2011年11月14日 月曜日 午後2時24分

Содержание

1 7301 TKE_nain book 32 2011 11 140 sea T Î 2HÇ24 6 Handling Errors and Problems 6 1 Error Messages Error Display iü ee os Cause Remedy Movement during measurement Repeat measurement Remain still and do not talk during measurement Refer to section 3 3 Air plug disconnected Insert the air plug securely Refer to section 3 2 Arm cuff not applied correctly Apply the arm cuff correctly Refer to section 3 2 Clothing is interfering with the arm cuff Remove any clothing interfering with the arm cuff Refer to section 3 2 Air is leaking from the arm cuff Replace cuff with new one Refer to Chapter 8 The arm cuff was inflated above 299 mmHg when inflating the cuff manually Remove the arm cuff and take another measurement Refer to section 3 3 Battery power is low Replace all four AAA batteries with new ones Refer to section 2 1 E DB y ______ Bl nksorappe KS continuously 32