6.2 TENS programmes
Prog.
No.
Reasonable areas of
application
Indications
Possible
electrode
position
Cycle 1 Cycle 2 Cycle 3 Cycle 4
Cycle 5 Cycle 6 Cycle 7 Cycle 8
Width
(µs)
Fre-
quen-
cy (Hz)
On
Time
(sec.)
Off
Time
(sec.)
Width
(µs)
Fre-
quen-
cy (Hz)
On
Time
(sec.)
Off
Time
(sec.)
Width
(µs)
Fre-
quen-
cy (Hz)
On
Time
(sec.)
Off
Time
(sec.)
Width
(µs)
Fre-
quen-
cy (Hz)
On
Time
(sec.)
Off
Time
(sec.)
1 + 11 Neck pain, tension
headache
01, 02, 13 250 4 30 0 250 4 30 0 250 5 30 0 250 5 30 0
250 6 20 0 250 6 20 0 250 8 30 0 250 8 30 0
2 + 12 Back pain 03, 04, 05,
06, 15, 23
250 6 30 0 250 6 30 0 250 8 20 0 250 8 20 0
250 10 20 0 250 10 20 0
3 + 13 Shoulder pain 07, 14 250 2 10 0 250 4 8 0 250 6 6 0
4 + 14 Pain as a result of
rheumatoid
arthritis
See
note 250 60 20 0 250 70 20 0 250 80 30 0 250 80 30 0
5 + 15 Lumbar pain 22 250 80 20 0 250 80 20 0 250 75 4 0 250 10 20 0
250 70 4 0 250 65 4 0
6 + 16 Menstrual pain 08 250 40 30 0 250 45 30 0 250 55 30 0 250 60 30 0
7 + 17 Pain programme I See note 250 4 30 0 250 4 20 0 250 6 30 0 250 6 20 0
250 8 30 0 250 8 20 0 250 10 30 0 250 10 20 0
8 + 18 Knee pain
Ankle joint pain,
capsular
injuries
09,
10 250 40 5 0 250 6 10 0 250 50 5 0
9 + 19
(Burst)
Pain programme II See note 250 75 0,25 0,25 250 2 0,5 0
10
+ 20
(Burst)
Pain programme III See note 250 100 0,25 0,25
On-Time (sec.) = Cycle switch-on time in seconds (contraction) – Off-Time (sec.) = Cycle switch-off time in seconds (relaxation)
Note: Electrode position is supposed to surround the painful area. With painful muscle groups, the electrodes are grouped round the muscle affected. In case of joint pain, the
joint on the front/rear side of the joint and, if electrode distances permit, on the right and left sides of the joint, should be surrounded with electrodes.
Minimal electrode distance should not be less than 5 cm and not exceed 15 cm. See figure no. 9 an 10 for knee and ankle joint.
Burst programmes are suitable for all sites which should be treated with an alternating signal pattern (to ensure minimum habituation).