Shimadzu UX-420H Руководство пользователя онлайн [86/129] 685321

Shimadzu UX-420H Руководство пользователя онлайн [86/129] 685321
70
14. Application Functions
Read Chapter 7 Before Use of These Pages
Notes
•Press the [POWER] key in the peak detection standby state to initiate
the power standby state.
•Press the [POWER] key during detection of the peak to return to the
peak detection standby state.
Peak Hold setting can be cleared without menu operation, by pressing
the [POWER] key for about three seconds.
Polarity of the peak value displayed is “polarity of the displayed value of
the first change (by five times or more of Zero Range) from the display
within Zero Range.”
Usually the peak value is easily measured by selecting menu item
(Pouring mode). Depending on the weighting conditions and the sample
type, this setting may not always be suitable.
Pretaring Value (13.7) function cannot be used with Peak Hold function.
•See 13.5 for details of Zero Range.
•See 11.3 for details of stability detection.
Not more than one of these seven functions can be set at a time. Setting
the second one with the menu will automatically clear the previously set
function; Auto Print(13.3), Auto Zero(13.4), Peak Hold(14.3), Interval
Timer(14.4), Add-on mode(14.5), Animal Weighing mode(14.6), Formu-
lation mode(14.7).
When minimum display has been switched to 10d (See 9.2), the opti-
mized Zero Range for 1d display may cause failure in the operation.
Select optimal Zero Range value again when minimum display has been
changed (See 13.5 for details).
23

Содержание

Скачать