Shimadzu UX-420H [87/129] Interval timer

Shimadzu UW-2200H [87/129] Interval timer
71
14. Application Functions
Read Chapter 7 Before Use of These Pages
14.4 Interval Timer
Not applicable to a verified balance as a legal measuring instrument in the EU
Automatically outputs the displayed value at preset intervals. The “T” symbol (“T” of the Tare symbol) is
illuminated when the Interval Timer is activated.
The optional AKB-301 Application Keyboard is convenient when the interval is changed frequently.
1 Select menu item and set the output inter-
val (00:01 = 1 sec to 99:59 = 99 minutes 59 sec-
onds).
2 In the interval timer standby state when the “T”
and the Stand-by symbols are both illuminated,
press the [PRINT] key.
The first data is output. Data will be automatically
output at the set time intervals thereafter.
3 To stop output, press the [POWER] key.
The balance returns to the interval timer standby
state in step 1).
50
Notes
Use the [O/T] key to erase the tare or zero the balance at any time.
•Pressing the [POWER] key in the interval timer standby state brings the
power supply standby state.
Interval Timer setting can be cleared without menu operation, by press-
ing the [POWER] key for about three seconds. This does not clear the
set time interval value.
Using the interval timer function to record data over a long period may
cause data error due to balance drift.
Some instruments receiving the data may not operate normally if the set
time interval is short. To correct this, set the time interval to a longer
period. When the set time interval is short and the instrument connected
to DATA I/O is unknown, it is recommended to set the handshake to a
setting other than “H-tm” .
Not more than one of these seven functions can be set at a time. Setting
the second one with the menu will automatically clear the previously set
function; Auto Print(13.3), Auto Zero(13.4), Peak Hold(14.3), Interval
Timer(14.4), Add-on mode(14.5), Animal Weighing mode(14.6), Formu-
lation mode(14.7).
76