Shimadzu UX-2200H [55/129] Caution

Shimadzu UW-2200H [55/129] Caution
39
10. Calibration
Read Chapter 7 Before Use of These Pages
10. Calibration
10.1 What is calibration?
Calibration is required to accurately weigh items with an electronic balance. Calibration should be per-
formed:
When the location of the balance is changed, even within the same room.
When the room temperature changes considerably.
Periodically, according to the quality control plan of the user.
Terms used in this manual:
Span Calibration: Adjustment of the balance to specifications using two weight values;
zero and an appropriate value for the balance capacity.
Calibration Check: Comparing the current calibration weight reading to the calibration
weight reading after the last span calibration.
Calibration: Pertains to both span calibration and calibration check.
Caution
Never plug off the balance when the following messages are displayed.
“i-CAL x”, “i-tESt x”, “wAit”, “Abort”, “CAL E x”(“x” represents a number).
With UW series, displaced built-in weight may cause damage to the mech-
anism.