Shimadzu UX-2200H [85/129] Peak hold

Shimadzu UW-2200H [85/129] Peak hold
69
14. Application Functions
Read Chapter 7 Before Use of These Pages
14.3 Peak Hold
Not applicable to a verified balance as a legal measuring instrument in the EU
Detects the peak value of a fluctuating weight. “Peak value” is the highest or lowest value displayed in
the duration after it has changed beyond five times the Zero Range until the first stability detection after
that. The “P” symbol (“P” of Auto-Print symbol) is illuminated when the Peak Hold function is activated.
Select menu item in advance.
1 In the peak detection standby state with the “P”
symbol and the Stand-by symbols illuminated,
press the [O/T] key to tare the display.
2 Press the [PRINT] key.
The Stand-by symbol disappears and peak value
detection starts.
3 Upon stability is detected, the operation of peak
value detection is ceased.
The detected peak value is displayed with the “P”
symbol and Asterisk simultaneously illuminated.
The peak value is output to external devices if
connected. After the detection of the peak value,
the same display remains regardless of the load
change on the pan.
4 Press the [POWER] key for the next operation.
The balance returns to the peak detection
standby state described in the above step 1.