Shimadzu UX-2200H [93/129] Formulation mode

Shimadzu UW-2200H [93/129] Formulation mode
77
14. Application Functions
Read Chapter 7 Before Use of These Pages
14.7 Formulation Mode
This function is convenient for weighing in the components of a formulation. The mass of each compo-
nent is displayed and stored Every time a component is added and [PRINT] key is pressed, the mass
of that component is output through RS-232C or DATA I/O interface and the display will be automati-
cally zeroed. When all the components have been weighed, the masses are summed up and the total
mass is displayed and output by pressing [POWER] key.
Select menu item .
(Operation)
1 When Formulation Mode is ON, the display
stands at Formulation stand-by status upon enter-
ing the mass display. Add-on symbol, Memory
symbol and Stand-by symbol are illuminated in
the display.
Place a container (if used) and press [O/T] key to
tare the display.
Note that taring with [O/T] key is not accepted
after [PRINT] key is pressed in the following step
2 until [POWER] key is pressed in the step 5.
2 Press [PRINT] key. When external device is
connected, “--------- FORMULATION MODE ---------”
will be output.
3 Load the first component and press [PRINT]
key. Upon stability is detected at a value higher
than five times Zero Range, the mass value will
be output with a numbering “CMP001 =”. The dis-
play will be automatically zeroed after that.
4 Repeat step 3 until all the components have
been weighed.
5 Press [POWER] key.
The total mass will be displayed and output to
external devices with a remark “TOTAL =”.
6 Clear the pan.
Start the next set of measurements from the step 1.
51
---Formulation Mode---
CMP001 =
0.5361g
CMP002 =
0.5422g
CMP003 =
0.4488g
TOTAL =
1.5271g
Output example: 1st component is 0.5361g,
2nd is 0.5422g, 3rd is 0.4488g.