Shimadzu UX-2200H Руководство пользователя онлайн [92/129] 685327

Shimadzu UX-2200H Руководство пользователя онлайн [92/129] 685327
76
14. Application Functions
Read Chapter 7 Before Use of These Pages
Notes
On the premise of weighing animated objects, the stability detection
band is automatically extended in the Animal Weighing mode. Repro-
ducibility of the measurement data is slightly less than with other modes.
If a larger stability detection band is selected with the menu (See 11.3.1),
stability detection is more readily made and the output of measurement
result usually becomes faster. However, the accuracy may become
poorer. When changing stability detection band, select according to the
specific measurement’s requirements on accuracy, observing the repro-
ducibility on the test run.
If the balance is slow to automatically return to the zero display, set a
larger Zero Range value.
Weighing unit cannot be changed when Animal Weighing is on.
•See 13.5 for details of Zero Range.
•See 11.3 for details of stability detection.
When Menu Lock (7.5.3) is activated, operational condition of Animal
Weighing is also locked.
When Menu reset (7.5.2) is performed, operational condition setting of
Animal Weighing is also returned to the default (Cond 1).
Pretaring Value (13.7) function cannot be used with Animal Weighing
mode.
Small animal bucket set (option, see A-3.) is recommended for weighing
small animals such as mice
When the [POWER] key is pressed in the Animal Weighing standby
state, the power standby state is initiated.
Animal Weighing mode setting can be cleared without menu operation,
by pressing the [POWER] key for about three seconds.
Not more than one of these seven functions can be set at a time. Setting
the second one with the menu will automatically clear the previously set
function; Auto Print(13.3), Auto Zero(13.4), Peak Hold(14.3), Interval
Timer(14.4), Add-on mode(14.5), Animal Weighing mode(14.6), Formu-
lation mode(14.7).
When minimum display has been switched to 10d (See 9.2), the opti-
mized Zero Range for 1d display may cause failure in the operation (e.g.
Data is not automatically output as animal is placed). Select optimal
Zero Range value again when minimum display has been changed (See
13.5 for details).

Содержание

Скачать