40
10. Calibration
Read Chapter 7 Before Use of These Pages
10.2 Calibration Execution
10.2.1 Span Calibration Using the Built-in Weight (UW Series Only)
The balance is adjusted using the built-in calibration weight.
1 Verify that the balance is in mass display and
that the pan is empty.
2 Press the [CAL] key once. “i-CAL” is displayed.
(If “i-CAL” is not displayed, return to mass display
and select menu item .)
3 Press the [O/T] key.
After “i-CAL3”, “i-CAL2”, “i-CAL1”, “Set”, “CALEnd”
the mass display will appear indicating the com-
pletion of span calibration.
Notes
Setting before shipment is as the following:
UW series: Span calibration using the built-in weight
UX series: Span calibration using external weights
The type of calibration can be changed (See 10.3).
Calibration will not be performed when the weight on the pan is not near
zero, or the balance is not stable.
...