Microlife BP B3 AFIB [10/20] Disposal

Microlife BP B3 AFIB [10/20] Disposal
8
uniformity of measurements. Non-invasive blood pressure meters.
Verification Method» Confirmation of initial verification, an elec-
tronic version of the verification certificate, is on the website of the
Federal Information Fund for Ensuring Uniformity of Measure-
ments - www.fgis.gost.ru
Information on verification is on the company website -
www.microlife.ru
Disposal
Batteries and electronic devices must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.
11.Guarantee
This device is covered by a 5 year guarantee from the date of
purchase. The guarantee is valid only on presentation of the guar-
antee card completed by the dealer (see back) confirming date of
purchase or the receipt.
Batteries and parts that become worn with use are not included.
Opening or altering the device invalidates the guarantee.
The guarantee does not cover damage caused by improper
handling, discharged batteries, accidents or non-compliance
with the operating instructions.
The cuff has a functional guarantee (bladder tightness) for
2
years.
Please contact your local Microlife-Service (see foreword).
12. Technical Specifications
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
Date of production: first four digits of the serial number of the
device. First and second digit – the week of the year / third and
fourth digit – the year of production.
Operating conditions:
10 - 40 °C / 50 - 104 °F
15 - 95 % relative maximum humidity
Storage conditions:
-20 - +55 °C / -4 - +131 °F
15 - 95 % relative maximum humidity
Weight:
402
g (including batteries)
Dimensions:
138 x 94.5 x 62.5
mm
Measuring procedure:
oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V diastolic
Measurement range:
20 - 280
mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure
display range:
0 - 299 mmHg
Resolution:
1 mmHg
Static accuracy:
pressure within ± 3 mmHg
Pulse accuracy:
± 5 % of the readout value
Voltage source:
4 x 1.5 V alkaline batteries; size AA
Mains adapter DC 6V, 600 mA (optional)
Battery lifetime: approx. 920 measurements
(using new batteries)
IP Class: IP20
Reference to
standards:
EN 1060-1 /-3 /-4; IEC 60601-1;
IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life: Device: 5 years or 10000 measurements
Accessories: 2 years

Содержание

Похожие устройства

Скачать