RTD Calibrator
Measure Temperature
17
Measure Temperature
Use Resistance-Temperature Detectors (RTDs)
The Product accepts RTD types shown in Table 6.
RTDs are characterized by their resistance at 0 °C (32 °F),
which is called the “ice point” or R
0
.
The most common R
0
is 100 Ω. The Product accepts RTD
measurement inputs in 2-, 3-, or 4-wire connections, with
the 3-wire connection the most common. A 4-wire
configuration provides the highest measurement precision,
and 2-wire provides the lowest measurement precision.
To measure temperature using an RTD input, proceed as
follows:
1. If necessary, push for Measure mode.
2. Push .
3. Use or to select the desired type, and push
to confirm.
4. Push
to select a 2-, 3-, or 4- wire connection.
5. Attach the RTD to input terminals as shown in
Figure 8.
Set Temperature Unit
The Product allows you to select the temperature unit °C
or °F.
Push to switch to shift mode, and push to set the
temperature unit to °C or °F.
The default temperature unit is °C.
Содержание
79- Users manual
- Rtd calibrator
- C users manual
- Warranty
- Table of contents
- C users manual
- Table title page
- List of tables
- List of figures
- Figure title page
- Introduction
- How to contact fluke
- C users manual
- Safety information
- Safe working practices
- Table 2 international electrical symbols
- C users manual
- Standard equipment
- Figure 1 standard equipment
- C users manual
- Input and output terminals
- C users manual
- The product has keys for different purposes some keys have secondary functions that are available when shift already shows on the display
- Rtd calibrator keys
- Figure 3 shows the product keys table 4 explains their use
- Figure 3 keys
- Figure 4 shows the elements of a typical display table 5 describes the elements
- Figure 4 elements of a typical display
- Display
- Rtd calibrator display
- C users manual
- Auto power off
- Auto backlight off
- Contrast adjustment
- C users manual
- Magnet mounting and hanging strap
- Measure ma current and activate loop power supply
- Measure ma current
- C users manual
- Activate loop power supply
- Use resistance temperature detectors rtds
- Set temperature unit
- Measure temperature
- C users manual
- Figure 8 measure temperature with an rtd
- Rtd calibrator measure temperature
- Source temperature signals
- C users manual
- Simulate rtds
- Figure 10 connections for simulating 3 and 4 wire rtd
- C users manual
- Set 0 and 100 output parameters
- Scale ma channel to temperature
- Step and ramp modes
- Set step and ramp range
- Select step or ramp mode
- C users manual
- Auto storage of settings
- Replace the batteries
- Clean the product
- C users manual
- Service center calibration or repair
- Maintenance
- Replacement parts
- Figure 12 replacement parts
- C users manual
- Specifications
- Ohms measurement
- Dc ma measurement
- Ohms source
- C users manual
- Year 2 year source current 1 year 2 year
- Rtd type α range c
- Rtd input and output
- Rtd calibrator specifications
- Measure c source c
- C users manual
- Loop power supply
- General specifications
Похожие устройства
-
Fluke 9144Руководство по эксплуатации -
Fluke 9143Руководство по эксплуатации -
Fluke 9142Руководство по эксплуатации -
Fluke 712CКраткий справочник -
Fluke 712CПриложение к Руководству по эксплуатации -
Fluke 712CПаспорт безопасности -
Fluke 712BРуководство по калибровке -
Fluke 712BКраткий справочник -
Fluke 712BРуководство по эксплуатации -
Fluke 712BПриложение к паспорту безопасности -
Fluke 712BПаспорт безопасности -
Fluke 712BПриложение к Руководству по эксплуатации
Узнайте, как использовать RTD калибратор для точного измерения температуры. Поддержка различных типов соединений и выбор единиц измерения.