Settings that change when Invert is set to On: .
The instrument adds a down arrow to the title row of the channel badge to indicate inverted mode
The channel Offset value reverses polarity
The trigger level value reverses polarity
The trigger level indicator moves to the inverted value and the trigger level bar appears momentarily to show the new level
The trigger Slope indicator on the Trigger badge and in its menu changes to the opposite edge (Rising to Falling, or Falling
to Rising). If the Either edge mode is selected then slope mode does not change
Settings that don't change when Invert is On:.
The channel Vertical scale does not change
The channel Vertical position knob rotation direction does not change
The channel Vertical Position value does not change
The Trigger level knob rotation direction does not change
Invert and Setup, Autoset, and Default functions.
The invert state is saved and recalled correctly as part of Setup and Session files
Autoset does not affect Invert
Default sets Invert to Off
Probe Setup panel (Channel configuration menu)
Use the Channel configuration menu Probe Setup panel to see probe information, check the probe compensation status,
compensate the probe, or restore the factory defaults.
To open the Probe Setup panel:
1. Double-tap an analog Channel badge on the Settings bar to open the Channel configuration menu.
2. Tap the Probe Setup panel.