Tektronix MSO46 [64/628] Powering on windows for the first time

Tektronix MSO56 [64/628] Powering on windows for the first time
4. Slide the connector end of the SSD into the drive connector bracket. Push firmly to seat the drive in the connector.
5. Push down on and tighten the thumb screw to attach the SSD drive to the chassis.
6. Reinstall the SSD drive cover on the bottom of the instrument.
7. Restore the instrument to its normal operating position.
8. Go to the Powering on Windows for the first time on page 34 procedure.
Powering on Windows for the first time
The instrument goes through a series of configuration bootups that require manual intervention when first powering up a new
(never-installed) Windows drive in an instrument. Once the drive is configured, the instrument boots into Windows and starts the
oscilloscope application from then on.
1. Power on the instrument. The oscilloscope powers up to initialize some settings, then shuts down.
2. Power on the instrument again. The instrument powers up into the application startup screen, then displays a message at
the bottom of the screen:
Updating System Files... This will take a few minutes.
The oscilloscope will shut down when the update is complete.
After a few minutes, the instrument powers down.
3. Power on the instrument again. The instrument boots up into Windows and starts the oscilloscope application.
The instrument attempts to activate the Windows license as part of the initial power-on process (See Activating Windows on
page 35).
Install your instrument
34 4/5/6 Series MSO Help

Содержание