Tektronix MSO46 — настройки триггера последовательного интерфейса RS-232 [486/628]

Tektronix MSO56 [486/628] Rs 232 serial bus trigger settings panel
RS-232 serial bus trigger settings panel. Requires option SRCOMP. Mode & Holdoff panel (Bus Trigger configuration panel)
fields and controls
Field or control Description
Source Select the RS232 bus on which to trigger.
Trigger On Select the type of information on which to trigger.
Data Bytes Sets the number of data bytes (1 byte = 8 bits) on which to trigger (one to ten bytes). Use the A
knob to change the value.
Available when Trigger On = Data.
Data Sets the data pattern on which to trigger. The number of bits shown depends on the Data
Words setting.
Tap the Binary, Hex, or ASCII field and use the A and B knobs to select and change the
values. Or double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Data.
A, B knob controls Use the A knob to select (highlight) the digit(s) to change.
Use the B knob to change the value of the digit(s).
Field or control Description
Trigger Mode The trigger mode determines how the instrument behaves in the absence or presence of a
trigger event:
Auto trigger mode enables the instrument to acquire and display a waveform even if a trigger
does not occur. Auto mode uses a timer that starts when the acquisition is started, and the
pretrigger information is obtained. If a trigger event is not detected before the timer times out,
the instrument forces a trigger. The length of time it waits for a trigger event depends on the
time base setting.
When forcing triggers in the absence of valid triggering events, Auto mode does not
synchronize the waveform on the display. The waveform will appear to jump across the screen.
If valid triggers occur, the display will become stable.
Normal trigger mode enables the instrument to acquire a waveform only when it is triggered. If
no trigger occurs, the last waveform record acquired remains on the display. If no last waveform
exists, no waveform is displayed.
Force Trigger Forces a trigger event regardless of whether the waveform meets any trigger conditions.