Fluke 1621 [17/26] Resistance limit mode
Превью страниц
Страница 17 /
26
![Fluke 1621 [17/26] Resistance limit mode](/views2/1004967/page17/bg11.png)
Earth Ground Tester
Features
9
Resistance Limit Mode
The Tester has a resistance limit mode that allows you to set a maximum
resistance reading. If a resistance reading exceeds the set limit, the Tester
beeps and the >LIMIT icon displays. The limit can be set between 0 and
1999 Ω.
To set the maximum resistance:
1. Set the rotary switch to LIMIT. If the limit mode is on, the Tester
displays the stored limit setting. If the limit mode is off, the Tester
displays OFF.
2. If the limit mode is off, press START. The Tester displays the stored
limit setting.
3. Press DISPLAY to step through the digits to select the digit that you
want to set. The first press of DISPLAY selects the leftmost digit.
(The digit flashes when selected.) The second and third press of
DISPLAY selects the second and third digits. The fourth press of
DISPLAY selects the decimal point to set the resolution of the
measurement.
4. With the digit selected that you want to set, press START to
increment the value. The leftmost digit increments from 0 to 19. The
other digits increment from 0 to 9. Or if you selected the decimal
point, press START to move the decimal point to change the
resolution of the measurement.
5. Repeat steps 3 and 4 until you have set the maximum resistance
value. When you have finished, set the rotary switch to OFF for 5
seconds to store the value.
Содержание
42- Earth ground tester
- Limited warranty and limitation of liability
- Title page
- Table of contents
- Table title page
- List of tables
- List of figures
- Figure title page
- Unpacking
- Packing
- Introduction
- Earth ground tester
- Xw warning
- Safety regulations
- Symbols
- Cat ii
- Accessories
- Features
- Software
- Lcd display
- To disable auto shut off simultaneously press and hold start and display while setting the rotary switch to any on position 3 pole 2 pole or limit to reset auto shut off or to reset the tester turn off the tester and then turn it on again
- The tester has an auto shut off feature that turns the device off after 10 minutes of inactivity the tester beeps to alert you that it is about to turn off
- The tester automatically checks for interference voltage noise fault voltage above 20 v interference above 20 v greatly diminishes the accuracy of measurements if the tester detects interference voltage above 20 v the measurement automatically stops the w icon flashes the u icon displays and the value of the stray voltage displays
- Interference detection
- Auto shut off
- Resistance limit mode
- Battery installation
- Operating instructions
- Xw warning
- Pole measurement
- Users manual
- Figure 4 three pole measurement setup
- Ac resistance measurement
- Troubleshooting
- To troubleshoot your tester follow the steps in table 4
- Specifications
- Measuring time 8 seconds average from when start is pressed
- Limit input tester retains set value even if instrument is turned off assuming battery power supply is sufficient
- If tester detects stray interference voltage 20 v w is displayed and the measurement is not started
- Earth ground tester specifications
- Storage
- Service
Похожие устройства
-
Fluke 568Инструкция к устройству -
Fluke 566Руководство пользователя -
Fluke 561Инструкция по применению -
Fluke 724Руководство по калибровке -
Fluke 724Паспорт безопасности -
Fluke 705Информация о калибровке -
Fluke 705Loop Calibrators Data Sheet -
Fluke 705Паспорт безопасности -
Fluke 705Приложение к паспорту безопасности -
Fluke 705Приложение к Инструкции -
Fluke 705Дополнение к информации о калибровке -
Fluke 707 exПриложение к Руководству по эксплуатации