Earth Ground Tester
Operating Instructions
11
Operating Instructions
XW Warning
To avoid possible electric shock or personal injury, before
powering up and operating the device, carefully read and
follow all safety regulations described in “Safety
Regulations.”
3-Pole Measurement
To perform a 3-pole measurement:
1. Insert the probe and auxiliary electrode stakes into the soil as shown
in Figure 4. Ensure the probe stake is a minimum distance of 20 m
(64 ft) from the earth ground electrode. Ensure the auxiliary electrode
stake is a minimum distance of 20 m (64 ft) from the probe stake.
Position the auxiliary electrode stake so it forms a straight line with
the earth ground electrode and probe stake.
2. Set the rotary switch to OFF.
3. Install the test leads as shown in Figure 4. Connect the earth ground
electrode to jack E/C1. Connect the probe to jack S/P2. Connect the
auxiliary electrode to jack H/C2.
4. Set the rotary switch to 3 pole and press START. The ACTIVE icon
displays to indicate that the measurement is in progress.
When the measurement is finished, the earth ground electrode
resistance (R
E) automatically displays. To display the auxiliary
electrode resistance (R
H), press DISPLAY. To display the probe
resistance (R
S), press DISPLAY again.
Содержание
42- Earth ground tester
- Limited warranty and limitation of liability
- Title page
- Table of contents
- Table title page
- List of tables
- List of figures
- Figure title page
- Unpacking
- Packing
- Introduction
- Earth ground tester
- Xw warning
- Safety regulations
- Symbols
- Cat ii
- Accessories
- Features
- Software
- Lcd display
- To disable auto shut off simultaneously press and hold start and display while setting the rotary switch to any on position 3 pole 2 pole or limit to reset auto shut off or to reset the tester turn off the tester and then turn it on again
- The tester has an auto shut off feature that turns the device off after 10 minutes of inactivity the tester beeps to alert you that it is about to turn off
- The tester automatically checks for interference voltage noise fault voltage above 20 v interference above 20 v greatly diminishes the accuracy of measurements if the tester detects interference voltage above 20 v the measurement automatically stops the w icon flashes the u icon displays and the value of the stray voltage displays
- Interference detection
- Auto shut off
- Resistance limit mode
- Battery installation
- Operating instructions
- Xw warning
- Pole measurement
- Users manual
- Figure 4 three pole measurement setup
- Ac resistance measurement
- Troubleshooting
- To troubleshoot your tester follow the steps in table 4
- Specifications
- Measuring time 8 seconds average from when start is pressed
- Limit input tester retains set value even if instrument is turned off assuming battery power supply is sufficient
- If tester detects stray interference voltage 20 v w is displayed and the measurement is not started
- Earth ground tester specifications
- Storage
- Service
Похожие устройства
-
Fluke 568Инструкция к устройству -
Fluke 566Руководство пользователя -
Fluke 561Инструкция по применению -
Fluke 724Руководство по калибровке -
Fluke 724Паспорт безопасности -
Fluke 705Информация о калибровке -
Fluke 705Loop Calibrators Data Sheet -
Fluke 705Паспорт безопасности -
Fluke 705Приложение к паспорту безопасности -
Fluke 705Приложение к Инструкции -
Fluke 705Дополнение к информации о калибровке -
Fluke 707 exПриложение к Руководству по эксплуатации
Узнайте, как правильно использовать тестер заземления с пошаговыми инструкциями. Соблюдайте правила безопасности и проводите измерения правильно.