Fluke 1621 — руководство пользователя: автоматическое переключение разрешения [26/26]
Превью страниц
Страница 26 /
26
![Fluke 1621 [26/26] Storage](/views2/1004967/page26/bg1a.png)
1621
Users Manual
18
Automatic changeover of resolution:
R
H Resolution
<7 kΩ 0.01 Ω
<50 kΩ 0.1 Ω
>50 kΩ 1 Ω
Interference Voltage Display DC + AC
Vmax: 30 Veff
Common mode
rejection:
>80 dB at 50 Hz and 60 Hz
Ri: 680 kΩ
Measuring uncertainty: <10 % for pure ac and dc signals
Storage
If the Tester is not going to be used or is being stored for a long period,
remove the battery and store separately from the Tester to avoid damage from
battery leakage.
Service
If you suspect that the Tester has failed, review this manual to ensure you are
operating it correctly. If the Tester still fails to operate properly, pack securely
using the original packing material (if available) and forward, postage paid, to
the nearest Fluke Service Center. Include a brief description of the problem.
Fluke assumes no responsibility for damage in transit.
To locate an authorized service center, call Fluke using the phone numbers
listed below:
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402 675 200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
You may visit us on the World Wide Web at www.fluke.com
. To register your
product, visit http://register.fluke.com
.
Содержание
42- Earth ground tester
- Limited warranty and limitation of liability
- Title page
- Table of contents
- Table title page
- List of tables
- List of figures
- Figure title page
- Unpacking
- Packing
- Introduction
- Earth ground tester
- Xw warning
- Safety regulations
- Symbols
- Cat ii
- Accessories
- Features
- Software
- Lcd display
- To disable auto shut off simultaneously press and hold start and display while setting the rotary switch to any on position 3 pole 2 pole or limit to reset auto shut off or to reset the tester turn off the tester and then turn it on again
- The tester has an auto shut off feature that turns the device off after 10 minutes of inactivity the tester beeps to alert you that it is about to turn off
- The tester automatically checks for interference voltage noise fault voltage above 20 v interference above 20 v greatly diminishes the accuracy of measurements if the tester detects interference voltage above 20 v the measurement automatically stops the w icon flashes the u icon displays and the value of the stray voltage displays
- Interference detection
- Auto shut off
- Resistance limit mode
- Battery installation
- Operating instructions
- Xw warning
- Pole measurement
- Users manual
- Figure 4 three pole measurement setup
- Ac resistance measurement
- Troubleshooting
- To troubleshoot your tester follow the steps in table 4
- Specifications
- Measuring time 8 seconds average from when start is pressed
- Limit input tester retains set value even if instrument is turned off assuming battery power supply is sufficient
- If tester detects stray interference voltage 20 v w is displayed and the measurement is not started
- Earth ground tester specifications
- Storage
- Service
Похожие устройства
-
Fluke 568Инструкция к устройству -
Fluke 566Руководство пользователя -
Fluke 561Инструкция по применению -
Fluke 724Руководство по калибровке -
Fluke 724Паспорт безопасности -
Fluke 705Информация о калибровке -
Fluke 705Loop Calibrators Data Sheet -
Fluke 705Паспорт безопасности -
Fluke 705Приложение к паспорту безопасности -
Fluke 705Приложение к Инструкции -
Fluke 705Дополнение к информации о калибровке -
Fluke 707 exПриложение к Руководству по эксплуатации
Изучите функции автоматического переключения разрешения и советы по обслуживанию. Узнайте, как правильно хранить и использовать устройство для предотвращения повреждений.