Fluke 707 [15/16] Accuracy specifications
English Instruction SheetPanel 15
Accuracy Specifications
Accuracy is specified for 1 year after calibration at operating
temperatures of 18 °C to +28 °C and is given as:
±( [ % of reading ]+[ counts ] )
MEASURE V dc
Range: +28 V (+30 V max)
Resolution: 1 mV
Input Impedance: 1 MΩ
Accuracy: ±(0.015 % of reading+2 counts)
MEASURE mA dc
Range: 20 mA (24 mA max)
Resolution: 1 μA
Accuracy: ±(0.015 % of reading+2 counts)
SOURCE / SIMULATE mA dc
Range: 0 mA to 20 mA (24 mA max)
Resolution: 1 μA
Accuracy: ±(0.015 % of reading+2 counts)
Source mode:
Compliance: To 1200 Ω at 20 mA
To 950 Ω at 20 mA in HART
TM
Mode
Simulate mode:
External loop voltage requirement: 24 V nominal, 30 V maximum,
12 V minimum
Loop Power
≥24 V
Percent display
-25 % to 125 %
Input / Output Protection
Fuseless protection
Содержание
- Battery saver 1
- Instruction sheet 1
- Introduction 1
- Loop calibrator 1
- English instruction sheetpanel 2 2
- Symbols 2
- W warnings and cautions 2
- Press in and tur 3
- Changing the ma output span 4
- English instruction sheetpanel 4 4
- Sourcing ma 4
- Using the ma sourcing output modes 4
- Contacting fluke 5
- English instruction sheetpanel 5 5
- A 12 v to 28 v loop supply must be available insert the test leads as shown below 6
- English instruction sheetpanel 6 6
- Simulating a transmitter 6
- When simulating the operation of a transmitter the calibrator regulates the loop current to a known value selected by you 6
- Auto ramping the ma output 7
- English instruction sheetpanel 7 7
- Using the spancheck function 7
- English instruction sheetpanel 8 8
- Measuring dc ma 8
- English instruction sheetpanel 9 9
- Measuring dc ma with loop power 9
- English instruction sheetpanel 10 10
- Measuring dc volts 10
- English instruction sheetpanel 11 11
- For maintenance procedures not described in this instruction sheet contact a fluke service center 11
- Maintenance 11
- W warning 11
- Calibration 12
- Cleaning 12
- English instruction sheetpanel 12 12
- In case of difficulty 12
- English instruction sheetpanel 13 13
- Hart mode 13
- Replacing the battery 13
- W warning 13
- English instruction sheetpanel 14 14
- Replaceable parts 14
- Accuracy specifications 15
- English instruction sheetpanel 15 15
- English instruction sheetpanel 16 16
- General specifications 16
Похожие устройства
- LG MH-6646QMS Инструкция по эксплуатации
- Supra SNP-433 Инструкция по эксплуатации
- Digitech XMM METAL MASTER Инструкция по эксплуатации
- Hotpoint-Ariston ARSD 129 Инструкция по эксплуатации
- Ada 5D Crosspoint Set А00192 Инструкция по эксплуатации
- Fluke 707 ex Инструкция по эксплуатации
- Supra SNP-431 Инструкция по эксплуатации
- Digitech BNX3 Инструкция по эксплуатации
- LG MH-6646QMB Инструкция по эксплуатации
- Hotpoint-Ariston ARUSL 105 Инструкция по эксплуатации
- Geo-Fennel FL 200 A-N 291000 Инструкция по эксплуатации
- Fluke 712 Инструкция по эксплуатации
- Digitech BP 200 Инструкция по эксплуатации
- Supra SNB-200 Инструкция по эксплуатации
- LG MH-6646QM Инструкция по эксплуатации
- Hotpoint-Ariston ARSL 105 (CSI).L Инструкция по эксплуатации
- Geo-Fennel FL 55 Multilinner 500050 Инструкция по эксплуатации
- Fluke 714 Инструкция по эксплуатации
- Digitech GNX2 Инструкция по эксплуатации
- Supra SNP-351 Инструкция по эксплуатации
Скачать
Случайные обсуждения