Documenting Process Calibrator
Using Source Mode
61
7. Press the
Ramp softkey. The display changes to
the following:
gj41s.eps
8. Fill in the parameters as prompted. Enter the
Start Value, End Value, and Ramp Time.
9. To automatically stop ramping if a trip condition
is detected, set the
Trip Detect to Enabled, and
select
Voltage or Continuity as the trip function.
10. Press the
Done softkey. Note the RAMP
annunciator next to SOURCE at the top of the
display.
11. Select a low-to-high ramp or a high-to-low ramp
with the
Ramp Up/Down softkey.
12. To start ramping, press the
Start Ramp softkey.
13. Ramping continues until a trip is detected (if
enabled), the ramp time expires, or you press
the
Stop Ramp softkey.