Microlife BP A90 [8/153] Using rechargeable batteries

Microlife BP A90 [8/153] Using rechargeable batteries
6
Using rechargeable batteries
You can also operate this instrument using rechargeable batteries.
6. Error Messages
If an error occurs during the measurement, the measurement is inter-
rupted and an error message, e.g. «ERR 3», is displayed.
)
Only use «NiMH» type reusable batteries.
)
Batteries must be removed and recharged when the flat
battery symbol appears. They should not remain inside the
instrument as they may become damaged (total discharged as
a result of low use of the instrument, even when switched off).
)
Always remove the rechargeable batteries if you do not intend
to use the instrument for a week or more.
)
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.
Error Description Potential cause and remedy
«
ERR 1
» Signal too
weak
The pulse signals on the cuff are too weak.
Re-position the cuff and repeat the
measurement.*
«
ERR 2
» Error signal During the measurement, error signals were
detected by the cuff, caused for instance by
movement or muscle tension. Repeat the
measurement, keeping your arm still.
«
ERR 3
» No pressure
in the cuff
An adequate pressure cannot be generated
in the cuff. A leak may have occurred. Check
that the cuff is correctly connected and is
not too loose. Replace the batteries if neces-
sary. Repeat the measurement.
«
ERR 5
» Abnormal
result
The measuring signals are inaccurate and no
result can therefore be displayed. Read
through the checklist for performing reliable
measurements and then repeat the
measurement.*
«
HI
» Pulse or cuff
pressure too
high
The pressure in the cuff is too high (over
300 mmHg) OR the pulse is too high (over
200 beats per minute). Relax for 5 minutes
and repeat the measurement.*
«
LO
» Pulse too low The pulse is too low (less than 40 beats per
minute). Repeat the measurement.*
Error Description Potential cause and remedy

Содержание

Using rechargeable batteries You can also operate this instrument using rechargeable batteries Error Description ERR 2 Error signal Potential cause and remedy During the measurement error signals were t Only use NiMH type reusable batteries detected by the cuff caused for instance by c movement or muscle tension Repeat the measurement keeping your arm still Batteries must be removed and recharged when the flat battery symbol appears They should not remain inside the instrument as they may become damaged total discharged as ERR 3 No pressure a result of low use of the instrument even when switched off c in the cuff Always remove the rechargeable batteries if you do not intend not too loose Replace the batteries if neces sary Repeat the measurement t Batteries cannot be charged in the blood pressure monitor ERR 5 Abnormal information regarding charging care and durability resuit The measu ring signal s are inaccurate an d n o result can therefore be displayed Read through the checklist for performing reliable 6 Error Messages measurements and then repeat the If an error occurs during the measurement the measurement is inter rupted and an error message e g ERR 3 is displayed measurement HI Error Description Potential cause and remedy ERR 1 Signal too The pulse signals on the cuff are too weak weak in the cuff A leak may have occurred Check that the cuff is correctly connected and is to use the instrument for a week or more Becharge batteries in an external charger and observe the An adequate pressure cannot be generated Re position the cuff and repeat the measurement Puise or cuff pressure too The pressure in the cuff is too high over 300 mmHg OR the pulse is too high over high 200 beats per minute Relax for 5 minutes and repeat the measurement LO Puise too low The pulse is too low less than 40 beats per minute Repeat the measurement 6 micro ife