Insulation Testers
Making Measurements
13
Measuring Insulation Resistance
Insulation tests should only be performed on de-energized
circuits. To measure insulation resistance set up the
Tester as shown in Figure 7 and follow the steps below:
1. Insert test probes in the
V and COM input terminals.
2. Turn the rotary switch to the desired test voltage.
3. Connect the probes to the circuit to be measured. The
Tester automatically detects if the circuit is energized.
• The primary display shows - - - - until you press
T and a valid insulation resistance reading
is obtained.
• The high voltage symbol (Z) along with a primary
display of >30 V warns if voltage more than 30 V
ac or dc is present. In this condition, the test is
inhibited. Disconnect the Tester and remove
power before proceeding.
4. Push and hold T to start the test. The secondary
display shows the test voltage applied to the circuit
under test. The high voltage symbol (Z) along with a
primary display showing the resistance in MΩ or GΩ
appears. The
t icon appears on the lower portion of
the display until T is released.
When resistance is higher than the maximum display
range, the Tester displays the Q symbol and the
maximum resistance for the range.
5. Keep the probes on the test points and release the
T button. The circuit under test then discharges
through the Tester. The resistance reading appears
on the primary display until a new test is started or a
different function or range is selected or > 30 V is
detected.
Содержание
58- 1503 insulation testers users manual
- Users manual
- Insulation testers
- Limited warranty and limitation of liability
- Limited warranty and limitation of liabilit
- To register your product online visit register fluke com
- Title page
- Table of contents
- Table title page
- List of tables
- List of figures
- Figure title page
- Contacting fluke
- Introduction
- Insulation testers
- Xw warning
- Safety information
- Rotary switch positions
- Battery saver sleep mode
- Unsafe voltage
- Buttons and indicators
- Xw warning
- Understanding the display
- Cal err
- Power up options
- Input terminals
- Making measurements
- Xw warning
- Measuring volts
- Measuring earth bond resistance
- Measuring insulation resistance
- Measuring polarization index and dielectric absorption ratios model 1507
- Using the compare function model 1507
- Testing the batteries
- Cleaning
- Xw warning
- Testing the fuse
- Xw warning
- Replacing the batteries and fuse
- 1503 users manual
- Specifications
- General specifications
- V 0 v 2 3
- Range resolution 50 hz to 400 hz
- Of rdg digits
- Insulation testers specifications
- Accuracy
- Ac dc voltage measurement
- Insulation specifications
- Earth bond resistance measurement
- Model 1507
- Model 1503
- En61557 specification
- Insulation resistance maximum and minimum display values
- Insulation resistance maximum and minimum display values cont
- Insulation resistance maximum and minimum display values cont
- Insulation resistance maximum and minimum display values cont
- Earth bond resistance maximum display values
Похожие устройства
-
Fluke 568Инструкция к устройству -
Fluke 566Руководство пользователя -
Fluke 561Инструкция по применению -
Fluke 724Руководство по калибровке -
Fluke 724Паспорт безопасности -
Fluke 705Информация о калибровке -
Fluke 705Loop Calibrators Data Sheet -
Fluke 705Паспорт безопасности -
Fluke 705Приложение к паспорту безопасности -
Fluke 705Приложение к Инструкции -
Fluke 705Дополнение к информации о калибровке -
Fluke 707 exПриложение к Руководству по эксплуатации