![Fluke 1507 [30/38] Insulation specifications](/views2/1004883/page30/bg1e.png)
1507/1503
Users Manual
22
Earth-bond Resistance Measurement
Range Resolution
Accuracy
1
+
(% of Rdg + Digits)
20.00 Ω 0.01 Ω
200.0 Ω 0.1 Ω
2000 Ω 1 Ω
20.00 kΩ 0.01 kΩ
+
(1.5 % + 3)
1. Accuracies apply from 0 to 100% of range.
Overload Protection ............................................2 V rms or dc
Open Circuit Test Voltage...................................> 4.0 V, < 8 V
Short Circuit Current ...........................................> 200.0 mA
Insulation Specifications
Measurement Range...........................................0.01 MΩ to 10 GΩ model 1507, 0.01 MΩ to 2000 MΩ model 1503
Test Voltages ......................................................50, 100, 250, 500, 1000 V model 1507, 500 and 1000 V model 1503
Test Voltage Accuracy ........................................+ 20 %, - 0 %
Short-Circuit Test Current ...................................1 mA nominal
Auto Discharge....................................................Discharge time <0.5 second for C = 1 µF or less
Live Circuit Detection: .........................................Inhibit test if terminal voltage > 30 V prior to initialization of test.
Maximum Capacitive Load..................................Operable with up to 1 µF load.
Содержание
58- 1503 insulation testers users manual
- Users manual
- Insulation testers
- Limited warranty and limitation of liability
- Limited warranty and limitation of liabilit
- To register your product online visit register fluke com
- Title page
- Table of contents
- Table title page
- List of tables
- List of figures
- Figure title page
- Contacting fluke
- Introduction
- Insulation testers
- Xw warning
- Safety information
- Rotary switch positions
- Battery saver sleep mode
- Unsafe voltage
- Buttons and indicators
- Xw warning
- Understanding the display
- Cal err
- Power up options
- Input terminals
- Making measurements
- Xw warning
- Measuring volts
- Measuring earth bond resistance
- Measuring insulation resistance
- Measuring polarization index and dielectric absorption ratios model 1507
- Using the compare function model 1507
- Testing the batteries
- Cleaning
- Xw warning
- Testing the fuse
- Xw warning
- Replacing the batteries and fuse
- 1503 users manual
- Specifications
- General specifications
- V 0 v 2 3
- Range resolution 50 hz to 400 hz
- Of rdg digits
- Insulation testers specifications
- Accuracy
- Ac dc voltage measurement
- Insulation specifications
- Earth bond resistance measurement
- Model 1507
- Model 1503
- En61557 specification
- Insulation resistance maximum and minimum display values
- Insulation resistance maximum and minimum display values cont
- Insulation resistance maximum and minimum display values cont
- Insulation resistance maximum and minimum display values cont
- Earth bond resistance maximum display values
Похожие устройства
-
Fluke 568Инструкция к устройству -
Fluke 566Руководство пользователя -
Fluke 561Инструкция по применению -
Fluke 724Руководство по калибровке -
Fluke 724Паспорт безопасности -
Fluke 705Информация о калибровке -
Fluke 705Loop Calibrators Data Sheet -
Fluke 705Паспорт безопасности -
Fluke 705Приложение к паспорту безопасности -
Fluke 705Приложение к Инструкции -
Fluke 705Дополнение к информации о калибровке -
Fluke 707 exПриложение к Руководству по эксплуатации
Изучите руководство пользователя для точного измерения сопротивления заземления. Узнайте о диапазонах, разрешении и точности измерений, а также о спецификациях изоляции.