Fluke 1507 — общие характеристики и спецификации устройства для измерений [28/38]
Превью страниц
Страница 28 /
38
![Fluke 1507 [28/38] 1503 users manual](/views2/1004883/page28/bg1c.png)
1507/1503
Users Manual
20
Specifications
Need intro paragraph her e
General Specifications
Maximum Voltage Applied to any Terminal ........600 V ac rms or dc
Storage Temperature..........................................-40 °C to 60 °C (-40 °F to 140 °F)
Operating Temperature.......................................-20 °C to 55 °C (-4 °F to 131 °F)
Temperature Coefficient......................................0.05 x (specified accuracy) per °C for temperatures < 18 °C or > 28 °C
(< 64 °F or > 82 °F)
Relative Humidity ................................................Noncondensing
0 % to 95 % @ 10 °C to 30 °C (50 °F to 86 °F)
0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F)
0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F)
Vibration..............................................................Random, 2 g, 5-500 Hz per MIL-PRF-28800F, Class 2 instrument
Shock ..................................................................1 meter drop per IEC 61010-1 2
nd
Edition (1 meter drop test, six sides, oak floor)
Electromagnetic Compatibility.............................In an RF field of 3 V/M, accuracy = specified accuracy (EN 61326-1:1997).
Safety..................................................................Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 NO. 61010-1-04, and
IEC/EN 61010-1 2
nd
Edition for measurement category IV 600 V (CAT IV)
Certifications .......................................................CSA per standard CSA/CAN C22.2 No. 61010.1-04; TUV per standard IEC/EN 61010-1
2
nd
Edition
Batteries..............................................................Four AA batteries (NEDA 15A or IEC LR6)
Battery Life..........................................................Insulation test use: Tester can perform at least 1000 insulation tests with fresh alkaline
batteries at room temperature. These are standard tests of 1000 V into 1 MΩ with a duty
cycle of 5 seconds on and 25 seconds off. Resistance Measurements: Tester can perform
at least 2500 earth-bond resistance measurements with fresh alkaline batteries at room
temperature. These are standard tests of 1 Ω with a duty cycle of 5 seconds on and 25
seconds off.
Size .....................................................................5.0 cm H x 10.0 cm W x 20.3 cm L (1.97 in H x 3.94 in W x 8.00 in L)
Weight.................................................................550 g (1.2 lb.)
Содержание
58- 1503 insulation testers users manual
- Users manual
- Insulation testers
- Limited warranty and limitation of liability
- Limited warranty and limitation of liabilit
- To register your product online visit register fluke com
- Title page
- Table of contents
- Table title page
- List of tables
- List of figures
- Figure title page
- Contacting fluke
- Introduction
- Insulation testers
- Xw warning
- Safety information
- Rotary switch positions
- Battery saver sleep mode
- Unsafe voltage
- Buttons and indicators
- Xw warning
- Understanding the display
- Cal err
- Power up options
- Input terminals
- Making measurements
- Xw warning
- Measuring volts
- Measuring earth bond resistance
- Measuring insulation resistance
- Measuring polarization index and dielectric absorption ratios model 1507
- Using the compare function model 1507
- Testing the batteries
- Cleaning
- Xw warning
- Testing the fuse
- Xw warning
- Replacing the batteries and fuse
- 1503 users manual
- Specifications
- General specifications
- V 0 v 2 3
- Range resolution 50 hz to 400 hz
- Of rdg digits
- Insulation testers specifications
- Accuracy
- Ac dc voltage measurement
- Insulation specifications
- Earth bond resistance measurement
- Model 1507
- Model 1503
- En61557 specification
- Insulation resistance maximum and minimum display values
- Insulation resistance maximum and minimum display values cont
- Insulation resistance maximum and minimum display values cont
- Insulation resistance maximum and minimum display values cont
- Earth bond resistance maximum display values
Похожие устройства
-
Fluke 568Инструкция к устройству -
Fluke 566Руководство пользователя -
Fluke 561Инструкция по применению -
Fluke 724Руководство по калибровке -
Fluke 724Паспорт безопасности -
Fluke 705Информация о калибровке -
Fluke 705Loop Calibrators Data Sheet -
Fluke 705Паспорт безопасности -
Fluke 705Приложение к паспорту безопасности -
Fluke 705Приложение к Инструкции -
Fluke 705Дополнение к информации о калибровке -
Fluke 707 exПриложение к Руководству по эксплуатации
Изучите основные характеристики и спецификации устройства для измерений, включая температурные диапазоны, влажность, вибрацию и безопасность. Подробности о батареях и сроке службы.