Fluke 1507 — how to Use Compare Function for Insulation Measurements [24/38]
Превью страниц
Страница 24 /
38
![Fluke FLK-1507 [24/38] Using the compare function model 1507](/views2/1004883/page24/bg18.png)
1507/1503
Users Manual
16
or
bbw10f.eps
Figure 8. Measuring Polarization Index and Dielectric
Absorption Ratios
Using the Compare Function (Model
1507)
Use the Compare function to set a pass/fail compare level
for the insulation measurements. To use the Compare
function:
1. Press the C button to select the desired compare
value. You can choose from 100 kΩ, 200 kΩ, 500 kΩ,
1 MΩ, 2 MΩ, 5 MΩ, 10 MΩ, 20 MΩ, 50 MΩ, 100 MΩ,
200 MΩ, and 500 MΩ.
2. Perform insulation tests as described earlier in this
manual.
3. The green pass indicator will appear if the measured
value is greater than the selected value.
4. Press and hold the C button for 1 second to
disable the Compare function. The pass indicator will
turn off when you start a new test or choose a new
compare value.
Содержание
58- 1503 insulation testers users manual
- Users manual
- Insulation testers
- Limited warranty and limitation of liability
- Limited warranty and limitation of liabilit
- To register your product online visit register fluke com
- Title page
- Table of contents
- Table title page
- List of tables
- List of figures
- Figure title page
- Contacting fluke
- Introduction
- Insulation testers
- Xw warning
- Safety information
- Rotary switch positions
- Battery saver sleep mode
- Unsafe voltage
- Buttons and indicators
- Xw warning
- Understanding the display
- Cal err
- Power up options
- Input terminals
- Making measurements
- Xw warning
- Measuring volts
- Measuring earth bond resistance
- Measuring insulation resistance
- Measuring polarization index and dielectric absorption ratios model 1507
- Using the compare function model 1507
- Testing the batteries
- Cleaning
- Xw warning
- Testing the fuse
- Xw warning
- Replacing the batteries and fuse
- 1503 users manual
- Specifications
- General specifications
- V 0 v 2 3
- Range resolution 50 hz to 400 hz
- Of rdg digits
- Insulation testers specifications
- Accuracy
- Ac dc voltage measurement
- Insulation specifications
- Earth bond resistance measurement
- Model 1507
- Model 1503
- En61557 specification
- Insulation resistance maximum and minimum display values
- Insulation resistance maximum and minimum display values cont
- Insulation resistance maximum and minimum display values cont
- Insulation resistance maximum and minimum display values cont
- Earth bond resistance maximum display values
Похожие устройства
-
Fluke 568Инструкция к устройству -
Fluke 566Руководство пользователя -
Fluke 561Инструкция по применению -
Fluke 724Руководство по калибровке -
Fluke 724Паспорт безопасности -
Fluke 705Информация о калибровке -
Fluke 705Loop Calibrators Data Sheet -
Fluke 705Паспорт безопасности -
Fluke 705Приложение к паспорту безопасности -
Fluke 705Приложение к Инструкции -
Fluke 705Дополнение к информации о калибровке -
Fluke 707 exПриложение к Руководству по эксплуатации
Learn how to use the Compare function to set pass/fail insulation measurement criteria. Set comparison levels for insulation resistance and monitor test results effectively.