IKA VACSTAR digital 0020016236 — цифровой вакуумный контроллер с регулировкой скорости [25/144]

IKA VACSTAR digital 0020016236 [25/144] Without
25
24 VDC
115/230 V
IKA Vacstar digital with vacuum controller VC 10
NOTE
Cannot be operated with optional
water valve
RV
10.5001 in conjunction
with VC 2.4.
Refer to the operating instructions
for the vacuum controller VC 10!
When the target value is reached, the suction line is closed by the
vacuum valve integrated in the VC 10. The pump speed can be
changed by manually setting the speed.
When using the optional pump cut-out unit
VC 2.4 (ref. 6) the pump is switched off when the target value is
reached (operating mode B).
This
device configuration
is also possible
without
the
pump
cut-out unit VC 2.4. The pump runs on. For good control, the
pump should be operated at a low speed.
1 Recipient (load, e.g. rotary evap-
orator, reactor)
2 Vacuum separator (e.g. Woulff
bottle)
3 Vent valve
4 Pressure sensor
5 Vacuum valve/ball valve
6 Pump cut-out unit VC 2.4
7 Emission condenser
Speed regulation
IKA Vacstar digital
with vacuum controller
VC 10 or
rotary
evaporator
RV 10 auto
NOTE
Observe the operating instructions
for the devices.
Automatc setting of the operating mode “Speed regulation” as
soon as the
analogue connecting
cable MVP
10.100 (ref. 8)
connects the pump with the vacuum controller
or
rotary
evaporator.
When the target
value
is reached, the
pump
speed is automatical-
ly reduced to a minimum of “0 rpm”. Once the target value is
reached, the pump applies suction according to the leakage rate
of the vacuum system. Automatic boiling point
recognition
is
possible.
1 Recipient (load, e.g. rotary evapo-
rator, reactor)
2 Vacuum separator (e.g. Woulff
bottle)
7 Emission condenser
8 Analogue connecting cable MVP
10.100