29
Error codes
If an error occurs, this is shown in an error on the display (B), e.g. Error 4.
Proceed as follows:
Switch the device off on the device switch,
Take corrective measure,
Switch device back on.
Error Cause Reason Solution
Error 3 Overheat The pump does not run Switch off the pump, disconnect it from the mains,
let it cool down.
Error 4 Motor blocked/overload The pump does not run Switch off the pump, disconnect it from the mains
Let the pump cool down.
Ventilate the vacuum system to normal pressure.
Error 8 Hall sensor defective The pump does not run Service
Error 9 BLP self-test/EEPROM error The pump does not run Service
No display Pump diaphragm defective Minimum pressure and suction
rate are not achieved
Service
Request a maintenance kit
No display Diaphragm valve defective Minimum pressure and suction
rate are not achieved
Service
Request a maintenance kit
No display Liquid in the pump chambers Minimum pressure and suction
rate are not achieved
Let the pump run dry (no-load operation)
In the event that the fault cannot be eliminated using the measures described or another error code is displayed:
Please contact the Service Department,
Send the device and a brief description of the fault.
Accessories
Vacuum controller IKA
VC 10
Vacuum control valve IKA
VCV 1 and VCV 2
Check valve IKA
VC 10.300
Vacuum hose IKA
VH.SI.8
PC 1.1 cable (RS 232)
Product contact parts
Name Material
Connection spigot PPS
Connection piece PPS
Distributor PPS
Head piece PPS
Diaphragm valve FFPM
Diaphragm NBR/PTFE
Clamp PPS
Maintenance kit
Vacuum safety emission condenser VSE 1
Woulff bottle/water valve set VSW1
For further accessories see www.ika.de