Electromagnetic Compatibility Information
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic emissions
The IH30 is intended for use in the electromagnetic environment specied below. The customer or the
user of the IH30 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1 The IH30 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emissions CISPR 11 Class B The IH30 is suitable for use in all establishments, inclu-
ding domestic establishments and those directly con-
nected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuations/icker
emissions IEC 61000-3-3
Complies
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic immunity
The IH30 is intended for use in the electromagnetic environment specied below. The customer or the
user of the IH30 should assure that it is used in such an environment.
IMMUNITY test IEC 60601 test level Compliance level Electromagnetic environment
– guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, con-
crete or ceramic tile. If oors
are covered with synthetic
material, the relative humidity
should be at least 30 %.
Electrical fast transi-
ent/burst
IEC 61000-4-4
±2 kV for power sup-
ply lines
±2 kV for power sup-
ply lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge IEC 61000-4-5
±1 kV line(s) and neu-
tral
±1 kV line(s) and neu-
tral
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
<5 % U
T
(>95 % dip
in U
T
) for 0,5 cycle 40
% U
T
(60 % dip in U
T
)
for 5 cycles 70 % U
T
(30 % dip in U
T
) for 25
cycles <5 % U
T
(>95
% dip in U
T
) for 5s
<5 % U
T
(>95 % dip
in U
T
) for 0,5 cycle 40
% U
T
(60 % dip in U
T
)
for 5 cycles 70 % U
T
(30 % dip in U
T
) for 25
cycles <5 % U
T
(>95 %
dip in U
T
) for 5s
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the IH30 requires continued
operation during power mains
interruptions, it is recommen-
ded that the IH30 be powered
from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz) magnetic
eld IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic
elds should be at levels cha-
racteristic of a typical location
in a typical commercial or hos-
pital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.