13. Enhancing Productivity
Read Chapter 7 Before Use of These Pages
Print on “GO” judge of checkweighing: Select menu item .
When the checkweighing function (See 13.1) is used and stability is detected in the “GO”
range, the data is output once.
The next data output is executed after the displayed value is within the Zero Range.
Note
During continuous output, the Communication symbol may appear to
remain lit. If the transfer speed of the data output is slow, the display may
become unstable. Increase the transfer speed as much as possible and
set the handshake off (menu item ).
Notes
•See 13.5 for details of Zero Range.
•See 11.3 for details of stability detection.
Pretaring Value (13.7) function cannot be used with Auto Print functions.
Auto Print mode setting can be cleared without menu operation, by
pressing the [POWER] key for about three seconds.
Not more than one of these seven functions can be set at a time. Setting
the second one with the menu will automatically clear the previously set
function; Auto Print(13.3), Auto Zero(13.4), Peak Hold(14.3), Interval
Timer(14.4), Add-on mode(14.5), Animal Weighing mode(14.6), Formu-
lation mode(14.7).
For automatic outputs at preset intervals, see “14.4 Interval Timer”.
When minimum display has been switched to 10d (See 9.2), the opti-
mized Zero Range for 1d display may cause failure in the operation (e.g.
Data is not automatically output as sample is loaded). Select optimal
Zero Range value again when minimum display has been changed (See
13.5 for details).