Multifunction Process Calibrator
Symbols
3
When servicing the Calibrator, use only specified replacement
parts. Do not open the Calibrator case. Opening the case
invalidates the Calibrator’s Ex Approval.
Do not allow water inside the case.
Before each use, verify the Calibrator’s operation by measuring a
known voltage.
Never touch the probe to a voltage source when the test leads are
plugged into the current terminals.
Do not operate the Calibrator around explosive dust.
When using a pressure module, make sure the process pressure
line is shut off and depressurized before connecting it or
disconnecting it from the pressure module.
Use four properly installed AA batteries to power the Calibrator.
Use only the batteries listed in Table 2.
Disconnect test leads from the circuit under test before changing
to another measure or source function.
Care must be taken to minimize the possibility of leakage when
measuring the pressure of toxic or flammable gases. Confirm that
all pressure connections are properly sealed.
Do not use in a damp or wet environment.
WCaution
To avoid possible damage to Calibrator or to equipment under test:
Disconnect the power and discharge all high-voltage capacitors
before testing resistance or continuity.
Use the proper jacks, function, and range for the measurement or
sourcing application.
Symbols
Symbols used on the Calibrator and in this manual are explained in Table 1.
Table 1. Symbols
Symbol Meaning Symbol Meaning
+ Power ON/OFF
T
Double insulated
J
Earth ground
)
Conforms to relevant Canadian and
US Standards. Certification #
LR110460-2
W
Risk of Danger. Important
information. Refer to manual.
P
Conforms to relevant European
Union directives.
M
Battery
F
Direct current
X
Hazardous Voltage
f
Pressure
(
Conforms to ATEX requirements ;
Conforms to relevant Australian
standards
~
Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for
recycling information.